Section 2 |
| |
|
|
|
Force Transmit Procedure
This diagnostic software routine allows the user to “force” a transmitting condition that can be use to search for an amplitude level (ALC) when Detection Fault or Low Signal alarms are present. The routine forces the flowmeter to generate constant transmit bursts while reporting current receive signal strength for the user. To initiate the Force Transmit function, refer to the example shown below.
1.After [Install] is selected press <ENT>. While the flowmeter is going through the drives (see menu screen below), press the <ALT> and <MENU> keys simultaneously.
| Siemens | 2 Channel | [1] SITE1 |
|
|
| Key [Install] after mounting transducers |
|
| ||
|
|
|
|
| |
| Transducer Model | 1011 Universal |
|
| |
| Transducer Size | B3 |
|
| |
| Xdcr Mount Mode | Direct |
|
| |
| Spacing Offset | Minimum |
|
| |
| Number Index |
| 4 |
|
|
| Spacing Method | Track 1012TP |
|
| |
| Ltn Value (in) |
| 0.217 |
|
|
| Install Completed? | No |
|
| |
| Empty Pipe Set | Channel Not Setup |
|
| |
| Zero Flow Adjust | Channel Not Setup |
|
| |
|
|
|
|
|
|
Pick/Install Xdcr
| Siemens | 2 Channel | [1] SITE1 |
|
|
| Drive 11 |
|
|
|
|
|
|
|
|
| |
| Transducer Model | 1011 Universal |
|
| |
| Transducer Size | B3 |
|
| |
| Xdcr Mount Mode | Direct |
|
| |
| Spacing Offset | Minimum |
|
| |
| Number Index |
| 4 |
|
|
| Spacing Method | Track 1012TP |
|
| |
| Ltn Value (in) |
| 0.217 |
|
|
| Install Completed? | Install |
|
| |
| Empty Pipe Set | Channel Not Setup |
|
| |
| Zero Flow Adjust | Channel Not Setup |
|
| |
|
|
|
|
|
|
Pick/Install Xdcr
NOTE: The <ALT> and <MENU> keys must be pressed before the meter scans through all the drives, or the selection of the detection mode and the Force Transmit function must be initiated again.
2.A typical menu screen will appear as shown below and indicate the current ALC (e.g., 50).
This ALC number indicates the current receive signal strength and can be used for further diagnos- tic purposes.
| Siemens | 2 Channel | [1] | SITE1 |
|
|
| ForceN | fx=8 | m=7 | ALC=50 |
|
|
|
|
|
|
| ||
| Transducer Model | 1011 Universal |
|
| ||
| Transducer Size | B3 |
|
|
| |
| Xdcr Mount Mode | Direct |
|
|
| |
| Spacing Offset | Minimum |
|
| ||
| Number Index | 4 |
|
|
| |
| Spacing Method | Track 1012TP |
|
| ||
| Ltn Value (in) | 0.217 |
|
|
| |
| Install Completed? | Install |
|
|
| |
| Empty Pipe Set | Channel Not Setup |
|
| ||
| Zero Flow Adjust | Channel Not Setup |
|
| ||
|
|
|
|
|
|
|
Pick/Install Xdcr
| Siemens | 2 Channel | [1] SITE1 | ||||
|
|
|
|
|
| ||
|
|
|
|
|
| ||
| Transducer Model | 1011 Universal |
| ||||
| Transducer Size | B3 |
| ||||
| Xdcr Mount | Mode |
|
|
| ||
| Spacing Offset | Detection Fault |
|
| |||
| Number Index |
| Press [ENT] |
|
| ||
| Spacing Method |
| 1012TP |
| |||
| Ltn Value (in) |
|
|
| 0.217 |
|
|
| Install Completed? | Install |
| ||||
| Empty Pipe Set |
| Channel Not Setup |
| |||
| Zero Flow Adjust | Channel Not Setup |
| ||||
|
|
|
|
|
|
|
|