Section 2

1010NFM-3J

 

 

 

Force Transmit Procedure

This diagnostic software routine allows the user to “force” a transmitting condition that can be use to search for an amplitude level (ALC) when Detection Fault or Low Signal alarms are present. The routine forces the flowmeter to generate constant transmit bursts while reporting current receive signal strength for the user. To initiate the Force Transmit function, refer to the example shown below.

1.After [Install] is selected press <ENT>. While the flowmeter is going through the drives (see menu screen below), press the <ALT> and <MENU> keys simultaneously.

 

Siemens

2 Channel

[1] SITE1

 

 

 

Key [Install] after mounting transducers

 

 

 

 

 

 

 

 

Transducer Model

1011 Universal

 

 

 

Transducer Size

B3

 

 

 

Xdcr Mount Mode

Direct

 

 

 

Spacing Offset

Minimum

 

 

 

Number Index

 

4

 

 

 

Spacing Method

Track 1012TP

 

 

 

Ltn Value (in)

 

0.217

 

 

 

Install Completed?

No

 

 

 

Empty Pipe Set

Channel Not Setup

 

 

 

Zero Flow Adjust

Channel Not Setup

 

 

 

 

 

 

 

 

Pick/Install Xdcr

 

Siemens

2 Channel

[1] SITE1

 

 

 

Drive 11

 

 

 

 

 

 

 

 

 

 

Transducer Model

1011 Universal

 

 

 

Transducer Size

B3

 

 

 

Xdcr Mount Mode

Direct

 

 

 

Spacing Offset

Minimum

 

 

 

Number Index

 

4

 

 

 

Spacing Method

Track 1012TP

 

 

 

Ltn Value (in)

 

0.217

 

 

 

Install Completed?

Install

 

 

 

Empty Pipe Set

Channel Not Setup

 

 

 

Zero Flow Adjust

Channel Not Setup

 

 

 

 

 

 

 

 

Pick/Install Xdcr

NOTE: The <ALT> and <MENU> keys must be pressed before the meter scans through all the drives, or the selection of the detection mode and the Force Transmit function must be initiated again.

2.A typical menu screen will appear as shown below and indicate the current ALC (e.g., 50).

This ALC number indicates the current receive signal strength and can be used for further diagnos- tic purposes.

 

Siemens

2 Channel

[1]

SITE1

 

 

 

ForceN

fx=8

m=7

ALC=50

 

 

 

 

 

 

 

 

Transducer Model

1011 Universal

 

 

 

Transducer Size

B3

 

 

 

 

Xdcr Mount Mode

Direct

 

 

 

 

Spacing Offset

Minimum

 

 

 

Number Index

4

 

 

 

 

Spacing Method

Track 1012TP

 

 

 

Ltn Value (in)

0.217

 

 

 

 

Install Completed?

Install

 

 

 

 

Empty Pipe Set

Channel Not Setup

 

 

 

Zero Flow Adjust

Channel Not Setup

 

 

 

 

 

 

 

 

 

Pick/Install Xdcr

 

Siemens

2 Channel

[1] SITE1

 

 

 

 

 

 

 

 

 

 

 

 

 

Transducer Model

1011 Universal

 

 

Transducer Size

B3

 

 

Xdcr Mount

Mode

 

 

 

 

Spacing Offset

Detection Fault

 

 

 

Number Index

 

Press [ENT]

 

 

 

Spacing Method

 

1012TP

 

 

Ltn Value (in)

 

 

 

0.217

 

 

 

Install Completed?

Install

 

 

Empty Pipe Set

 

Channel Not Setup

 

 

Zero Flow Adjust

Channel Not Setup

 

 

 

 

 

 

 

 

 

Pick/Install Xdcr

2-27