7.4MOUNTING THE TRANSDUCERS

Two mounting configurations are available. Adjacent mounting locates the two transducers alongside each other using a single mounting chain or strap. In-Line mounting locates the two transducers axially along the pipe using two mounting chains or straps. On metal pipes the Receive Transducer is always located behind the Transmit Transducer. On plastic pipes the Transmit Transducer is always located behind the Receive Transducer. In both cases, the cable entry will be on the same end of both trans- ducers unlike the transit-time configurations. Adjacent mounting will provide the maximum sensitivity to flow. Refer to the appropriate transducer installation drawing for the 191N1S (Installation Drawing 191N1S-7).

“C2”

C2 Transducers Transducers

Connect to UP Connect to UP

Terminal on flow terminal on flow

computercomputer

Flow Direction

Flow Direction

Connect to DOWN Connect to Terminal on flow DOWN terminal

computer on flow computer

REFLEXOR MODE - ADJACENT TRANSDUCER MOUNTING

“C2”C2 TransducersTransducers

Connect to Connect to DOWN DOWN Terminal terminal on flow on flow computer computer (RECEIVE) (Receive)

Connecton ct toto UPTerminalt r inalonon

flflow computercomputer(Transmit)(TRANSMIT)

“C2”

C2 Transducers

Transducers

Flow Direction

Flow Direction

Flowl Directionirection

NOTE: Transducer position shown is for metal pipes.

REFLEXOR MODE - IN-LINE TRANSDUCER MOUNTING

7-3