Section 5 | |
|
|
5.3PIPE CONSIDERATIONS FOR CLAMP-ON TRANSDUCER
5.3.1 PIPE DIMENSIONS
Controlotron manufactures transducer assemblies to service pipes from 0.25" to 360" in outer diam- eter. During the transducer install procedure, the flow computer will recommend transducer sizes based on the site data that you enter. Pipes with
5.3.2 PICKING THE APPROPRIATE TRANSDUCER
To ensure that you select the appropriate transducers for your application, consider the pipe outer diameter, temperature range and degree of precision required. Decide whether you need two indepen- dent flow channels or if you have to operate the system in
Initial transducer size recommendations are based on your pipe outer diameter, wall thickness and pipe material entries. However, you can override the meter recommendations to accommodate spe- cific situations. For example, you may own a set of transducers whose size does not appear on the meter list. You may be able to use transducers on a pipe size outside of their nominal pipe OD range. During transducer install, the flow computer recommends Spacing Indices (for spacer bars and mounting tracks) or Ltn, (actual distance required between transducers). Once you mount transducers and invoke the install routine, you can determine the suitability of your transducers by checking the signal strength and comparing the reported sonic velocity with the actual sonic velocity of the liquid to be measured.
NOTE: Transducer pairs must have matching serial numbers.
5.3.3 FLOW VELOCITY RANGE
System 1010’s flow velocity range with
5.3.4 OVERVIEW OF SYSTEM PERFORMANCE
Our system performance specifications are based on acceptable liquid sonic conductivity and other pertinent application conditions. The diversity that characterizes liquid flow, makes it impossible for us to cover all possible application conditions that have the potential to reduce performance. Performance