Section 2

1010NFM-3J

 

 

 

Assigning Vo Output Functions

The Vo analog output is a 0-10 VDC signal that varies linearly in relation to a selected function.

To assign a function to the voltage output:

To access the [Vo] option list press

To move the cursor to the desired data function press

To store selection press ENT

Assigning Pgen Output Functions

The Pgen analog output is a buffered TTL-compliant pulse rate signal, similar to the pulse outputs of flowmeters such as turbines. It varies linearly from 0-5000 Hz in relation to a selected data function.

To assign a function to the Pgen output:

To access the [Pgen] option list press

To move the cursor to the desired data function press

To store selection press ENT

PULSE OUTPUT (Pgen Wiring)

 

 

 

1010N & 1010DN with 1010N-2 Modules

TB2

 

Installation Drawing 1010N-2-7 (Sheet 2 of 2)

PIN#

SIGNAL

 

FUNCTION

NOTES

11

PGEN 1+

 

FREQUENCY OUTPUT 1

0 - 5000 Hz, 5 Volt Logic, Square Waveform

12

PGEN 1-

 

REF. GROUND

 

13

PGEN 2+

 

FREQUENCY OUTPUT 2

 

14

PGEN 2-

 

REF. GROUND

 

 

1010N & 1010DN with Expanded 1010N-7 I/O Modules

 

 

Installation Drawing 1010N-7-7 (Sheet 2 of 3)

TB2

(Standard, High Performance and Enhanced Performance Flowmeters)

 

 

 

 

PIN#

SIGNAL

 

FUNCTION

NOTES

9

PG1

 

FREQUENCY OUTPUT 1

0 - 5000 Hz, 5 Volt Logic, Square Waveform

10

PG2

 

REF. GROUND

 

11

PG3

 

FREQUENCY OUTPUT 2

 

12

PG4

 

REF. GROUND

 

2.10.2 RELAY SETUP

Use this menu to assign a function to channel relays. System 1010 supports 2 types of relay outputs, Alarm Relay and Pulse Relay. Alarm Relay outputs operate in “fail-safe” mode. The relay(s) are ener- gized under normal conditions - an alarm condition causes the relay(s) to de-energize until alarm clears. The Pulse Relay output supports totalizer and batch relay functions. Output is a pulse width approximately 200 ms; maximum activation rate is 2.5 pulses per sec. If totalizer pulses exceed this rate, excess pulses are stored in an overflow register. This allows the relay to “catch up” when flow decreases enough.