Section 1 |
| |
|
|
|
Siemens
NOTE: You are allowed to override the transducer type and size mounting mode and spacing offset. See paragraph 1.7.
Now you have to mount the transducers in accordance with the data recommended on the screen. Refer to the Section 3 for detailed instructions. The pipe must be completely filled with a liquid for the initial installation of the meter. Make sure that you do not install your transducers backwards or reverse the Upstream and Downstream cables.
Siemens
Use the [Install Completed?] menu cell to inform the meter that you have finished mounting the
hardware. Press. Move the cursor to [Install]. Press ENT
The meter performs an Initial Makeup which conditions its operation to the selected transducer, pipe and liquid and then sets the empty pipe and the zero flow levels automatically. Be patient! The Initial Makeup performs a number of operations including a frequency sweep of the pipe wall resonance, which could take several minutes.
Siemens