Siemens SITRANS FUS1010 NEMA-4X/ IP 65 Display of F AT no Flow Conditions, Other Menu Entries

Models: SITRANS FUS1010 NEMA-4X/ IP 65

1 232
Download 232 pages 51.98 Kb
Page 203
Image 203

Section 7

1010NFM-3J

 

 

 

 

Siemens

2 Channel [1]

Channel 1

 

Shows range of flow rates detected

 

 

 

 

 

 

Low Limit GAL/MIN

0.00

 

 

High Limit GAL/MIN

16.75

 

 

Noise Level Set

 

2

 

 

Doppler Freq <Hz>

8571

 

 

% Deviation

>

100.00

 

 

Signal Pk-Pk mV

27

 

 

FFTs/Sec

 

28

 

 

FFT Peak

 

0

 

 

Carrier fx <drive>

27

 

Application Info

7.14DISPLAY OF “F” AT NO FLOW CONDITIONS

It is normal for the digital display to show an “F” for the no flow condition since the lack of Spectra information during no flow is the same as a fault condition.

NOTE: Selecting the Operation Adjust menu cell [Zero/Fault Set] allows the option of indicating Zero Flow rather than a Fault (see paragraph 2.5.4 for details).

7.15SELECTION OF LIQUID COMPOSITION

The width of the Spectra wave shape is an indication of the depth of penetration of sonic energy into the flow stream. The % Deviation is a measure of the width of the Spectra and is representative of the penetration of the sonic beam into the liquid. Select [Slurry] in the Application Data menu for better flow rate indication accuracy when the % Deviation is less than 25 with typical flow conditions. For % Devia- tions above 35, select [Liquid]. For applications that continuously have % Deviations between 25 and 35, proper selection of either [Liquid] or [Slurry] can be confirmed by comparison to a known flow reference.

 

Siemens

2 Channel

[1]

Channel 1

 

 

 

Choose Liquid or Slurry flow calibration

 

 

 

 

 

 

 

 

 

 

 

 

Liquid Composition

Liquid

 

 

 

 

 

Density SG

 

1.00

 

 

 

 

 

 

 

 

 

 

 

Application Data

7.16OTHER MENU ENTRIES

All other menu entries will operate and be used in the same manner as the transit time flow measure- ment mode.

7-9

Page 203
Image 203
Siemens SITRANS FUS1010 NEMA-4X/ IP 65 manual Display of F AT no Flow Conditions, Selection of Liquid Composition