Section 2

1010NFM-3J

 

 

 

NOTE: Do not use ASA schedule code to specify the wall thickness. You must enter actual dimensions.

To enter the Pipe Wall Thickness:

To enable numeric entry press

Use the keypad’s numeric keys to type exact wall thickness (use in/mm).

To register the pipe wall thickness press ENT

2.2.6 LINER MATERIAL

For lined pipes, select a pipe liner from the material option list. If the pipe liner material does not appear on the list then select the closest type available. If necessary, call Siemens Customer Service for additional help. The system default Liner Material is [None]. If the pipe does not have a liner, press the <Down Arrow> twice to bypass to the next two menu cells. The [Liner Material] option list offers a selection of common liner materials.

To select a Liner Material:

To access the [Liner Material] option list press

To scroll to the required Liner Material press

To register selection press ENT

2.2.7 LINER THICKNESS

If you specified a pipe liner in the [Liner Material] menu cell, then use this menu cell to set its exact thickness in appropriate units (English or Metric).

To enter a Liner Thickness:

To enable numeric entry press

Use the number keys to type the exact line thickness value.

To register the data press ENT

2.3THE APPLICATION DATA MENU

This Menu becomes available after picking a Meter Type, Measurement Channel and Measurement Technology. Use of this menu is optional.

CAUTION: An erroneous Viscosity entry could affect the meter’s intrinsic calibration. Do not alter the default viscosity value of your liquid unless you are sure of your data.

Use the Application Data menu to edit default settings to match your application. When you specify a Liquid Class, the meter will adjust its operation to accommodate the liquid’s estimated sonic velocity, viscosity and specific gravity. If necessary, you can edit each liquid class parameter individually to obtain a closer match with the liquid. The default liquid is Water at 68°F (20°C).

2-10

Page 50
Image 50
Siemens SITRANS FUS1010 NEMA-4X/ IP 65 manual Liner Material, Liner Thickness, Application Data Menu