Section 2

 

 

 

 

1010NFM-3J

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

Siemens

Dual Path

4SS10G

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

fx (drive)

29

 

 

 

 

 

 

 

N (burst length)

5

 

 

 

 

 

 

 

Ltn in

5.316

 

 

 

 

 

 

 

Vf max CU

 

 

 

 

 

 

 

 

Vs max m/s

 

 

 

 

 

 

 

 

Vs min m/s

1355.00

 

 

 

 

 

 

 

Samples/Cycle

29

 

 

 

 

 

 

 

HF

>-0.120

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

Site Setup Data

 

 

 

 

 

 

 

 

 

 

 

 

 

 

“Automatic” Adjustment Procedure

zIn the [Site Setup Data] Menu, press the <Down Arrow> and scroll to the [HF] menu cell. Press the <Right Arrow> and a pop-up [Manual] prompt will appear.

zPress the <Up or Down Arrow> to select [Automatic] then press <ENT>.

zThe current measured correction value is displayed (see below).

zPress <ENT> again to install this correction value which will now appear next to the [HF] menu cell.

NOTE: The value shown in the [Automatic] pop-up prompt can not be changed and is for user information only.

Siemens

Dual Path

4SS10G

 

 

 

Siemens

Dual Path

4SS10G

 

 

 

fx (drive)

29

N (burst length)

5

Ltn in

 

5.316

 

Vf max CU

 

Adjustment

Vs max m/s

 

>Automatic

Vs min m/s

 

1355.00

 

Samples/Cycle

29

HF

>-0.000

fx (drive)

N (burst length) Ltn in

Vf max CU Vs max m/s Vs min m/s Samples/Cycle

HF

29

5

5.316

>Automatic

=0.026

1355.00

29

>-0.026

Site Setup Data

Site Setup Data

zIf you decide not to use the [Automatic] selection, press any key other than <ENT> to abort the operation.

2-71

Page 111
Image 111
Siemens SITRANS FUS1010 NEMA-4X/ IP 65 manual Automatic Adjustment Procedure