Section 2 |
|
|
|
| ||||
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
| Siemens | Dual Path | 4SS10G | |||
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
| fx (drive) | 29 |
|
|
|
|
|
|
| N (burst length) | 5 |
|
|
|
|
|
|
| Ltn in | 5.316 |
|
|
|
|
|
|
| Vf max CU |
|
|
|
|
|
|
|
| Vs max m/s |
|
|
|
|
|
|
|
| Vs min m/s | 1355.00 |
|
|
|
|
|
|
| Samples/Cycle | 29 |
|
|
|
|
|
|
| HF |
|
|
|
| |
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
| Site Setup Data |
|
|
|
|
|
|
|
|
|
|
|
|
|
|
“Automatic” Adjustment Procedure
zIn the [Site Setup Data] Menu, press the <Down Arrow> and scroll to the [HF] menu cell. Press the <Right Arrow> and a
zPress the <Up or Down Arrow> to select [Automatic] then press <ENT>.
zThe current measured correction value is displayed (see below).
zPress <ENT> again to install this correction value which will now appear next to the [HF] menu cell.
NOTE: The value shown in the [Automatic]
Siemens | Dual Path | 4SS10G |
|
|
|
Siemens | Dual Path | 4SS10G |
|
|
|
fx (drive) | 29 | |
N (burst length) | 5 | |
Ltn in |
| 5.316 |
| ||
Vf max CU |
| Adjustment |
Vs max m/s |
| >Automatic |
Vs min m/s |
| 1355.00 |
| ||
Samples/Cycle | 29 | |
HF |
fx (drive)
N (burst length) Ltn in
Vf max CU Vs max m/s Vs min m/s Samples/Cycle
HF
29
5
5.316
>Automatic
=0.026
1355.00
29
Site Setup Data
Site Setup Data
zIf you decide not to use the [Automatic] selection, press any key other than <ENT> to abort the operation.