RESET PROCEDURE FOR BLIND SYSTEMS
1010N Blind systems allow you to perform a system reset via the
NOTE: Custom
To Clear Active Memory Using the RS-232 Interface
1.Turn off power (if it is currently on). Turn power on. As soon as you apply power, immediately type the @ character three times. The prompt: [Clr Active Memory? No] appears at the top of the screen.
2.Press the <Right Arrow> and then the <Down Arrow> to switch the option list to:
[Clr Active Memory? Yes] Press <ENT> to clear all Active Site Data (but not saved site setups).
3.To restore operation, press <MENU> to access the Installation Menu. Create a new site setup or recall a stored site setup.
To Clear All Dynamic Memory Using the RS-232 Interface:
CAUTION: Before proceeding further , it is essential to understand that this function eliminates all data stored in RAM. This means that all saved site setups, including the site data of a
1.Turn off power (if it is currently on).
2.Turn the power on. As soon as you apply power, type the @ character three times.
The prompt: [Clr Active Memory?] appears at the top of the screen. Press the <Down Arrow>.
Note that the prompt switches to [Clr Saved Data? No] .
3.Press the <Right Arrow> and then the <Down Arrow> to switch the option list to: [Clr Saved Data? Yes].
4.Press <ENT> to clear all Saved Site Data, Datalogger Data, user created Pipe Data and Transducer Data.
5.To restore operation, press <MENU> to access the Installation Menu. Create a new site setup or recall a stored site setup. Reselect any Meter Facilities items (e.g.,