Siemens SITRANS FUS1010 NEMA-4X/ IP 65 manual HOW to Select a Transducer Size

Models: SITRANS FUS1010 NEMA-4X/ IP 65

1 232
Download 232 pages 51.98 Kb
Page 63
Image 63

Section 2

1010NFM-3J

 

 

 

To select a Transducer Type:

To access the [Xdcr Type] option list press

To move the cursor to the required transducer model press

To store your selection press ENT

2.4.2 HOW TO SELECT A TRANSDUCER SIZE

When you move the cursor to Transducer Size, the highlighted prompt at the top of the display screen shows a list of recommended transducer sizes. For example [Recommended Xdcrs: D3,D2,D1,C1,C3]. The flow computer generates this list automatically based on your pipe and your application data entries. The left-most transducer size (e.g., D3) is the primary (most appropriate) choice. The right-most transducer (e.g., C3) is still acceptable (but the least desirable) choice. Since these are simply recommendations not requirements, you can override the flow computer and use any size, as long as the mounting method accommodates the pipe’s diameter. The flow computer shows that your selection is acceptable if it is able to calculate transducer spacing (Ltn).

To select a Transducer Size:

To access the [Transducer Size] option list press

To move the cursor to the desired transducer size press

To select the transducer press ENT

2.4.3 HOW TO SELECT A TRANSDUCER MOUNT MODE

The flow computer recommends the transducer mounting mode [Direct or Reflect]. In almost all cases it will be Reflect, since this is the most desirable configuration. Reflect mode allows you to mount the transducers on pipes that do not permit back or side access. Reflect mode provides inherent compen- sation for flow profile distortion (crossflow) that could be the result of unfavorable application conditions such as an insufficient upstream straight run. In addition, Reflect mode supports the AutoZero function, which sets the zero flow velocity automatically.

Direct mounting may produce a stronger transducer signal and requires less mounting length than Reflect mounting. This can be important if the liquid or pipe material exhibits high sonic attenuation, thereby preventing operation in the Reflect configuration.

NOTE: We recommend using Direct Mount with Plastic pipes.

To select a Xdcr Mounting Mode:

To access the [Xdcr Mount] option list press

To move the cursor to the required mounting type press

To store selection press ENT