Section 6

1010NFM-3J

 

 

You can complete the Installation procedure without actually mounting the flow tube at its intended location. Fill the flow tube completely with liquid and connect the Up and DN cables according to the Installation/Outline Drawing. Complete the instructions in this section. If you choose to mount the Flow Tube first, please conform to the guidelines presented in this section and the Installation Drawing. In any case, do not proceed with these instructions until the flow tube is filled completely with liquid.

After filling the flow tube with liquid, pressto access the [Install Completed?] option list and then move the cursor to [Install] by pressing

 

Siemens

2 Channel

[1]

FLOW1

 

 

 

Key Install to start operation

 

 

 

 

 

 

 

 

 

 

 

 

Select FlowTube

1S

 

 

 

 

 

Install Completed?

>Yes

 

 

 

 

 

Empty FlowTube Set

New Makeup

 

 

 

 

Zero Flow Adjust

Chan Not Setup

 

 

 

 

 

 

 

 

 

 

Install FlowTube

To invoke install routine press ENT . This triggers Initial Makeup (an internal process that interrogates the flow tube and analyzes setup data to optimize operation). During procedure, the meter shows its progress by highlighting the prompt line.

 

Siemens

2 Channel [1]

FLOW1

 

Drive 27

 

m 10 [6:

:0]

 

 

 

 

 

 

 

Select FlowTube

1S

 

 

 

Install Completed?

Install

 

 

 

Empty FlowTube Set

Chan Not Setup

 

 

Zero Flow Adjust

Chan Not Setup

 

 

 

 

 

 

 

Install FlowTube

Upon completion, the computer reports the measured sonic velocity. You can accept this reading or

press to adjust it (see field manual for instructions).

Note that the default Empty Flow Tube Set is [MTYmatic]. The flow computer deter- mines the empty flow tube setting automatically during the Initial Makeup, so use of this menu cell is optional.

6-7

Page 189
Image 189