Section 5

 

 

1010NFM-3J

 

 

 

 

 

 

 

 

 

 

 

 

 

 

RECOMMENDED SONIC COUPLING COMPOUNDS

 

 

 

ITEM

CC#

USE

CHARACTERISTICS

 

 

1

101

Temporary and portable use on Clamp-on

Low in halogens and

 

 

 

 

 

flow transducers. Temp. range: -40°F to 100°F.

sulfur content.

 

 

2

102

Temporary and portable use for all Clamp-on flow

Low in halogens and sulfur

 

 

 

 

 

transducers. Well suited for survey use.

content. Water soluble.

 

 

 

 

 

Temp. range: -30°F to 100°F.

 

 

 

 

3

109

Temporary and portable use for all Clamp-on

Low viscosity Petroleum

 

 

 

 

 

flow transducers. Temp. range: -40°F to 300°F.

product.

 

 

4

110

High temp. temporary couplant to be used when

Resists breakdown and

 

 

 

 

 

setting up high temp. plastic flow transducers.

thinning out at high temper-

 

 

 

 

 

Temp. range: 255°F to 450°F.

atures.

 

5

111

Permanent high temp. couplant for plastic flow

Room temperature vulcanizing

 

 

 

 

 

transducers. Temp. range: 255°F to 450°F.

silicone rubber.

 

6

112

Permanent ambient temperature couplant for

Room temperature vulcanizing

 

 

 

 

 

 

plastic flow transducers. Temp. range: -40°F

silicone rubber.

 

 

 

 

 

to 250°F.

 

 

 

 

7

113

Protective spray for providing corrosion resistance

Spray on liquid. Excellent

 

 

 

 

 

to exposed transducer and track parts.

resistance to salt water

 

 

 

 

 

 

atmosphere.

 

8

114

Temporary and long term couplant for all flow

Teflon filled grease.

 

 

 

 

 

transducer types. Temp. range: -40°F to 255°F.

Water resistant.

 

9

117

Long term and permanent couplant for Clamp-On

Silicone grease loaded with

 

 

 

 

 

 

transducers. Temp. range: -40°F to 450°F.

metal oxides. Resists Break-

 

 

 

 

 

 

down and thinning at high

 

 

 

 

 

 

temperatures.

 

10

120

Submersible couplant. To be used on all submer-

Extremely water resistant.

 

 

 

 

 

ged transducer applications.

Consult Controlotron before

 

 

 

 

 

 

using.

 

11

122

Long term and permanent couplant for Clamp-

Long chain polymer grease.

 

 

 

 

 

 

on transducers. Temp. range: -40°F to 450°F.

Resists breakdown and thin-

 

 

 

 

 

 

ning at high temperatures.

 

12

124

Long term and permanent couplant for Clamp-On

Two part room temperature

 

 

 

 

 

transducers. Temp. range: -40°F to 250°F.

cure adhesive.

 

 

13

128

Temporary and long term couplant for all flow

High viscosity silicone

 

 

 

 

 

 

transducer types. Temp. range: -40°F to 375°F.

grease. Water resistant.

 

 

14

129

Adhesive backed Dry/Damping couplant for all flow

Viscoelastic polymer sheet.

 

 

 

 

129A

transducer types.

Water resistant.

Notice: Siemens holds US Patent Number 4,929,368 on the CC#122 Coupling Compound and all compounds of this type. Its use is restricted to Siemens products only, unless a special license has been obtained. Licenses for use with other products are available through the Siemens Sales Department.

5-9

Page 174
Image 174
Siemens SITRANS FUS1010 NEMA-4X/ IP 65 manual Recommended Sonic Coupling Compounds, CC# USE Characteristics