Siemens SITRANS FUS1010 NEMA-4X/ IP 65 NFM-3J Low Valc, Detection Fault/Low Signal, High Vaer

Models: SITRANS FUS1010 NEMA-4X/ IP 65

1 232
Download 232 pages 51.98 Kb
Page 127
Image 127

Section 2

1010NFM-3J

 

 

 

Low Valc

Clean and condition pipe surface.

Use High Precision or first recommended Universal Transducer.

Use Direct Mount (be sure of transducer 180º alignment!).

Use larger size transducer (Universal Only!).

Use wider spacing (Maximum or Letter D).

Stop flow to check for Aeration/Cavitation.

Try another location.

Perform an Initial Makeup after ANY of the above changes!!

Detection Fault/Low Signal

Perform all checks for “Low Valc”

Ensure pipe is full.

Ensure cables are terminated at both ends; Check continuity with Ohmmeter. Check operation on Pipe Simulator.

Switch off any radiating electrical equipment in area (Transmitters, Variable Freq. Drives, etc.).

High Vaer

Avoid transducer mounting near flow obstructions.

Check for possible pressure drop sources (Pipe Expansion, Orifice, Check Valve, etc.). Check liquid “Flash Point” against application conditions.

Pinch downstream valve (of Transducer) to raise pressure.

Mount on suction side of pump.

Raise tank levels; adjust Float switches.

Switch off any radiating electrical equipment in area (Transmitters, Variable Freq. Drives, etc.).

Poor Signal

Check that actual pipe dimensions and material match programmed data.

Consult X-Select; Choose first recommended transducer.

Relocate transducers.

2-87

Page 127
Image 127
Siemens SITRANS FUS1010 NEMA-4X/ IP 65 manual NFM-3J Low Valc, Detection Fault/Low Signal, High Vaer, Poor Signal