Section 3

1010NFM-3J

3.1.8USING 1012T MOUNTING TRACKS

The 1012TP and 1012THP Mounting Tracks provide a rigid mounting platform for Series 1011 universal or high precision size A or B transducers. The mounting tracks service pipe sizes up to a maximum of 5.00” (140 mm) outer diameter. Operating temperatures are supported up to 250°F (121°C). The assembly consists of lightweight aluminum track rails with integral transducer clamping screws. Attached index pins enable positive locating of the transducers at fixed spacing locations. Roller-chains and tension screws secure the assembly to the pipe. The following instructions refer to 1011 universal transducers. Please refer to Engineering Drawing 1012TP-7 for reflect mounting of high precision transducers and 1012THP-7 for direct mounting of high precision transducers.

The 1012T mounting tracks support both Direct and Reflect mounting modes. The flow computer recommends the appropriate transducers, mounting track and mounting mode, based on the pipe data entries. Refer to the instructions in paragraph 2.4 for details on the Transducer Installation proce- dure. If necessary, review paragraphs 3.1.2 through 3.1.4 for details on how to select and prepare a mounting location on your pipe.

Installing a 1012T Mounting Track in Reflect Mode

Paragraph 2.4 describes the Transducer Installation procedures that lead up to the automatic selection of transducers, mounting mode and spacing method. Examine the figure below, which shows a typical Pick/Install Xdcr menu screen. Note the automatic assignment of model numbers for the transducer and mounting track, plus the designation of the number index.

 

Siemens

2 Channel

[1] SITE1

 

Key [Install] after mounting transducers

 

 

 

 

 

 

Transducer Model

1011 Universal

 

 

Transducer Size

B3

 

 

Xdcr Mount Mode

Reflect

 

 

Spacing Offset

 

Minimum

 

 

Number Index

 

6

 

 

 

 

 

 

 

 

 

 

 

Spacing Method

Track 1012TP

 

 

Ltn Value (in)

 

0.581

 

 

 

 

Install Completed?

No

 

 

 

 

Empty Pipe Set

 

Channel Not Setup

 

 

 

 

Zero Flow Adjust

Channel Not Setup

 

 

 

 

 

 

 

 

 

Pick/Install Xdcr

Note selection of type, size and mounting mode.

Note selection of mounting track part number and number index.

This example requires a Model 1012T Mounting Track to accommodate size B3 universal transduc- ers. Note the reported number index. You will be inserting an index pin into this hole on the track rail to position one of the transducers (see diagram on next page).

1.Perform all required menu steps up until the point where you respond to the [Install Completed?] prompt. Note the reported number index. You will be inserting index pins into this hole and the reference hole on the track rail (see diagram on next page). Check to ensure that you have a matched set of transducers. They both should have the same S/N number but marked with either an “A” or “B” (e.g., 100A and 100B).

3-11

Page 139
Image 139