Section 1 |
Upon completion, the meter reports the measured liquid sonic velocity. Press. The flowmeter will then automatically perform an AutoZero operation, where the meter measures any zero flow offset, which is used to correct the flow reading. This operation occurs only when the transducers are mounted in the reflect configuration and is independent of whether or not the liquid is flowing.
After the AutoZero operation is complete, the meter may ask you to register the transducer used in this installation. It is advisable that you register your transducers, since the registration list is used to condition the transducer recommendation.
Press and select [YES], or [NO] if you do not want to register this transducer and then, press ENT .
Siemens
The meter is now ready to report flow. Press MENU. The meter will then ask you to save the Site. Saving the Site is optional and will therefore be covered in paragraph 1.7. To abort this item,
press . This brings up a Flow Display Screen.
Siemens | 2 Channel [1] | Channel 1 |
0FLOW GAL/MIN.2089.68
TOTAL KGAL | Vs | Vfo |
0.12 | 1474 | 0.00 |
| G/M | |
Once your site has successfully installed, you can press | to scroll between the avail- | |
able display screens. |
|
|
The meter is now operational and the display will show flow by pressing the <MENU> key.
Now that you are familiar with the 1010N setup, take the time to read the information about the Installa- tion Menu in Section 2. Becoming acquainted with the various flowmeter setups in the Main Menu will help further your understanding of the 1010N.