Siemens SITRANS FUS1010 NEMA-4X/ IP 65 Accuracy, Repeatability, Data Stability, Data Scatter

Models: SITRANS FUS1010 NEMA-4X/ IP 65

1 232
Download 232 pages 51.98 Kb
Page 168
Image 168

Section 5

1010NFM-3J

 

 

within specifications depends primarily on the receive signal’s signal-to-noise ratio and amplitude. The information below may point to application conditions that could reduce system performance below its normally high level.

5.3.5 ACCURACY

Although system accuracy is exceptional over a wide turndown ratio, at extremely low flow rates, a small zero offset becomes a high percentage of actual flow. Obviously, the ultimate accuracy will be obtained by performing an on-site flow calibration. A flow calibration can increase system accuracy to between 0.3% to 0.5%, depending on application conditions.

Two common data-entry mistakes may reduce performance. If you enter an incorrect liquid viscosity value, you could compromise the intrinsic flow profile compensation curve. Incorrectly identifying the transducers will reduce accuracy. Measured sonic velocity (Vs) errors will usually reveal this problem and by simply returning to the appropriate menu cells, entering the correct values, and then repeating the transducer installation it will resolve it.

5.3.6 REPEATABILITY

Some applications require repeatability rather than absolute accuracy. System 1010 features excellent repeatability specifications since its digital “no moving parts” design avoids the adverse effects of hysteresis and other wear mechanisms typical of mechanical devices.

5.3.7 DATA STABILITY

Two main factors influence the system’s data stability: Data Scatter and Drift.

Data Scatter

Data scatter is a rapid variation in flow readings (within a span of about 0.1 to 5 seconds). Minimal data scatter (approximately 0.01 to 0.03 ft/sec) is a natural by-product of digital computation that extracts the extremely small difference in the up vs. down sonic transit time. Minimal data scatter will not influ- ence the integrated flow total over periods as short as several minutes. Naturally, it will be a greater percentage of the reading when the meter measures extremely low flow rates. Poor liquid sonic con- ductivity may attenuate sonic signal to a level that increases data scatter. You should check the signal level (Valc %) item on the Diagnostic Menu. Usually, this is indicated by a low Valc % value (less than 30).

System 1010 does not exhibit inertia since it has no moving parts. In addition, it takes readings ten times per second. Therefore, it can detect and track very brief flow fluctuations that are beyond the response capability of some conventional meters. This performance level is required for detecting very fast and short flow transients or for a fast-response servo control loop application. However, you can use the damping and slewing controls to smooth the output response if you want the system to ignore rapid flow fluctuations or data scatter.

Data Drift

Drift is a defined as a long-term cyclical flow deviation resulting from the variation of liquid temperature or liquid sonic velocity. Drift may be more noticeable when combined with a poor signal-to-noise ratio. System 1010 is carefully designed to minimize the effects of drift. There are no drift-prone analog phase-locked loop devices in the primary detection circuits. In addition, we use only the most stable plastics or steel to construct our transducers.

5-3

Page 168
Image 168
Siemens SITRANS FUS1010 NEMA-4X/ IP 65 manual Accuracy, Repeatability, Data Stability, Data Scatter, Data Drift