Monitoring and Operating the Controller

RUN/HOLD key

You can start or stop the Ramp or Program using the RUN/HOLD key.

4.19Setpoint Ramp

Introduction

When you have configured a SETPOINT RAMP, the ramp will occur between the current local setpoint and a final local setpoint over a time interval of from 1 to 255 minutes. You can RUN or HOLD the ramp at any time.

Configuration Check

Make sure

SPRAMP is enabled

A Ramp Time (TIMIN) in minutes has been configured

A final setpoint value (FINLSP) has been configured. See Subsection 3.5 – Configuration group “SPRAMP” for details.

Operation

Running a Setpoint Ramp includes starting, holding, viewing the ramp, ending the ramp and disabling it. See Table 4-24.

Table 4-24 Running A Setpoint Ramp

Step

Operation

Press

Result

1Select Automatic Mode

M-A

Reset

“A” indicator is on.

Upper Display = Hold and PV value Lower Display = SP and Present value

2Set Start Setpoint

3Start the Ramp

Lower

Display

Run

Hold

Until start SP value is in lower display

Upper Display = Hold and PV value

Lower Display = SP and start SP value

You will see

Upper Display = Run and a changing PV value

Lower Display = SP and a changing SP value increasing or decreasing toward a final SP value

4Hold/Run the Ramp

5View the remaining ramp time

Run

Hold

Lower

Display

This holds the ramp at the current setpoint value. Press again to continue.

Until you see

Upper Display = RUN or HOLD and the PV value Lower Display = RP xx HH.MM (time remaining)

116

UDC2500 Universal Digital Controller Product Manual

4/07

Page 130
Image 130
Honeywell UDC2500 manual 19Setpoint Ramp, RUN/HOLD key, Running a Setpoint Ramp