100 Chapter 2
Performance Verification Tests
13. Input Attenuation Switching Uncertainty

13. Input Attenuation Switching Uncertainty

A 50 MHz CW signal is applied to the input of the analyzer through two
calibrated step attenuators. The attenuators are the amplitude
reference standard. The source is adjusted for a response at the
reference level. The internal attenuators are then varied between
settings and the external attenuators are changed accordingly to
maintain the same input level at the mixer. The analyzer marker
functions are used to measure the amplitude differences. The actual
attenuation values of the step attenuators are used to correct the
marker amplitude readings yielding the input attenuation switching
error.
The related adjustment for this performance test is “Frequency
Response.”
Equipment Required
Synthesized signal generator
1 dB step attenuator
10 dB step attenuator
Attenuator switch driver (if programmable step attenuators are
used)
6 dB fixed attenuator (2 required)
Cable, Type-N 152-cm (60-in) (2 required)
Cable, BNC 122-cm (48-in)
Attenuator interconnect kit
Additional Equipment for 75 Input
50 to 75 minimum loss pad
Adapter, Type-N (f), to BNC (m), 75
Additional Equipment for Option BAB
Adapter, Type-N (f), to APC 3.5 (f)