Chapter  3 607
Performance Verification Test Records
Agilent E4408B Performance Verification Test Record
60 dB –0.70 dB (12)________  0.70 dB ±0.100 dB
65 dB –0.75 dB (13)________  0.75 dB ±0.111 dB
15. Reference Level 
Accuracy
Log
     Reference Level
–10 dBm –0.3 dB (1)________  0.3 dB ±0.091 dB
 0 dBm –0.3 dB (2)________  0.3 dB ±0.100 dB
–30 dBm –0.3 dB (3)________  0.3 dB ±0.092 dB
–40 dBm –0.3 dB (4)________  0.3 dB ±0.094 dB
–50 dBm –0.5 dB (5)________  0.5 dB ±0.095 dB
–60 dBm –0.5 dB (6)________  0.5 dB ±0.097 dB
–70 dBm –0.5 dB (7)________  0.5 dB ±0.099 dB
–80 dBm –0.7 dB (8)________  0.7 dB ±0.101 dB
Linear
     Reference Level
–10 dBm –0.3 dB (9)________  0.3 dB ±0.091 dB
 0 dBm –0.3 dB (10)________  0.3 dB ±0.100 dB
–30 dBm –0.3 dB (11)________  0.3 dB ±0.092 dB
–40 dBm –0.3 dB (12)________  0.3 dB ±0.094 dB
–50 dBm –0.5 dB (13)________  0.5 dB ±0.095 dB
–60 dBm –0.5 dB (14)________  0.5 dB ±0.097 dB
–70 dBm –0.5 dB (15)________  0.5 dB ±0.099 dB
–80 dBm –0.7 dB (16)________  0.7 dB ±0.101 dB
16. Resolution Bandwidth 
Switching Uncertainty
Table 3-14  Agilent E4408B Performance Verification Test Record
Agilent Technologies
Model E4408B Report No. ___________ 
Serial No. ___________ Date ___________
Test Description Minimum Results 
Measured Maximum Measurement 
Uncertainty