
340 Chapter 2 
Performance Verification Tests
42. Tracking Generator Absolute Amplitude and Vernier Accuracy: Agilent E4402B, E4403B, 
E4404B, E4405B, E4407B, and E4408B (Option 1DN)
42. Tracking Generator Absolute Amplitude
and Vernier Accuracy: Agilent E4402B,
E4403B, E4404B, E4405B, E4407B, and E4408B
(Option 1DN)
A calibrated power sensor is connected to the tracking generator output 
to measure the power level at 50 MHz. 
The power meter is set to relative mode so that future power level 
readings are in dB relative to the reference power level setting. The 
output power level setting is decreased in 1 dB steps and the power 
level is measured at each step. The difference between the ideal and 
actual power levels is calculated at each step. 
Since a power sweep is accomplished by stepping through the vernier 
settings, the peak-to-peak variation of the vernier accuracy is equal to 
the power sweep accuracy. 
The related adjustment for this performance test is the front panel 
Align Now, TG (Ext Cable) under the System, Alignments menu.
Equipment Required 
Power meter, compatible with power sensor
Power sensor, 50 Ω 
Procedure 
1. Press Preset on the analyzer. Press the Factory Preset softkey, if it is 
displayed. Set the analyzer by pressing the following keys: 
FREQUENCY, Center Freq, 50 MHz
SPAN, Zero Span
AMPLITUDE, 0 dBm
System, Alignments, Auto Align, Off 
Source Amptd, Amplitude (On), –20 dBm 
Source Amptd, Attenuation (Man), 16 dB
Single Sweep
2. Zero and calibrate the power meter and power sensor in log mode 
(power reads out in dBm), as described in the power meter operation 
manual. Enter the 50 MHz Cal Factor of the power sensor into the 
power meter. 
3. Connect the 50 Ω power sensor to the RF output as shown in Figure 
2-60.
Figure 2-60  Absolute Amplitude, Vernier, and Power Sweep Accuracy Test