464 Chapter 3
Performance Verification Test Records
Agilent E4402B Performance Verification Test Record
−40 dBm Reference Level
–40 dBm input –0.54 dB (11)________ 0.54 dB ±0.087 dB
–50 dBm input –0.54 dB (12)________ 0.54 dB ±0.087 dB
−50 dBm Reference Level
–50 dBm input –0.54 dB (13)________ 0.54 dB ±0.087 dB
21. Resolution Bandwidth
Accuracy
Resolution Bandwidth
5 MHz 3.5 MHz (1)________ 6.5 MHz ±110 kHz
3 MHz 2.55 MHz (2)________ 3.45 MHz ±66 kHz
1 MHz 0.85 MHz (3)________ 1.15 MHz ±22 kHz
300 kHz 255 kHz (4)________ 345 kHz ±6.6 kHz
100 kHz 85 kHz (5)________ 115 kHz ±2.2 kHz
30 kHz 25.5 kHz (6)________ 34.5 kHz ±660 Hz
10 kHz 8.5 kHz (7)________ 11.5 kHz ±220 Hz
3 kHz 2.55 kHz (8)________ 3.45 kHz ±66 Hz
1 kHz 850 Hz (9)________ 1.15 kHz ±22 Hz
120 kHz 96 kHz (10)________ 144 kHz ±2.17 kHz
9 kHz 7.2 kHz (11)________ 10.8 kHz ±163 Hz
23. Frequency Response Note: Enter data in the appropriate section below depending upon
the ambient temperature at which the test was performed and
installed options.
20 to 30° C
Non-Option UKB
Band 0
(9 kHz to 3.0 GHz)
Maximum Response (1)________ 0.46 dB ±0.19 dB
Table 3-4 Agilent E4402B Performance Verification Test Record
Agilent Technologies
Model E4402B Report No. ___________
Serial No. ___________ Date ___________
Test Description Minimum Results
Measured Maximum Measurement
Uncertainty