Chapter  3 429
Performance Verification Test Records
Agilent E4401B Performance Verification Test Record
50 Ω Termination  ___________ ___________ ___________
Minimum Loss Pad
(Option 1DP only) ___________ ___________ ___________
1 dB Step Attenuator ___________ ___________ ___________
10 dB Step Attenuator ___________ ___________ ___________
Oscilloscope
(Option 1D6 only) ___________ ___________ ___________
Microwave Spectrum
Analyzer (Option 1DN
or 1DQ only) ___________ ___________ ___________
Notes/comments: _____________________________________________________________ 
_____________________________________________________________
Table 3-2  Agilent E4401B Performance Verification Test Record
Agilent Technologies
Model E4401B Report No. ___________ 
Serial No. ___________ Date ___________
Test Description Minimum Results 
Measured Maximum Measurement 
Uncertainty
1. 10 MHz Reference 
Output Accuracy 
(Non-Option 1D5 only)
Settability –5.0 Hz (1)________ 5.0 Hz ±293.3 µHz
2. 10 MHz High-Stability 
Frequency Reference 
Output Accuracy 
(Option 1D5 only)
5 Minute Warm-Up Error –0.1 ppm (1)________ 0.1 ppm ±0.000072 ppm
15 Minute Warm-Up Error –0.01 ppm (2)________ 0.01 ppm ±0.000070 ppm
3. Frequency Readout and 
Marker Frequency 
Count Accuracy
Frequency Readout 
Accuracy
 Center Freq     Span 
   1490 MHz      20 MHz 1489.83 MHz (1)________ 1490.17 MHz ±0 Hz