452 Chapter  3 
Performance Verification Test Records
Agilent E4402B Performance Verification Test Record
  Center Freq    Span
    1500 MHz     10 MHz 1499.999999 
MHz
(4)________ 1500.000001 
MHz
±0 Hz
    1500 MHz     1 MHz 1499.999999 
MHz
(5)________ 1500.000001 
MHz
±0 Hz
6. Frequency Span 
Accuracy 
 Span        Start Freq 
 3000 MHz    0 Hz  2370 MHz (1)________ 2430 MHz ±6.12 MHz
 100 MHz     10 MHz 79 MHz (2)________ 81 MHz ±204 kHz
 100 kHz     10 MHz 79 kHz (3)________ 81 kHz ±204 Hz
 100 MHz     800 MHz 79 MHz (4)________ 81 MHz ±204 kHz
 100 kHz     800 MHz 79 kHz (5)________ 81 kHz ±204 Hz
 100 MHz     1400 MHz 79 MHz (6)________ 81 MHz ±204 kHz
 100 kHz     1499 MHz 79 kHz (7)________ 81 kHz ±204 Hz
7. Noise Sidebands
 Offset from 1 GHz signal 
 (Serial Prefix < US4510, 
SG4510 or MY4510)
       10 kHz  (1)________ –90 dBc/Hz ±2.44 dB
       20 kHz (2)________ –100 dBc/Hz ±2.44 dB
       30 kHz (3)________ –106 dBc/Hz ±2.44 dB
 (Serial Prefix ≥ US4510, 
SG4510 or MY4510, 
Option 1DR)
       10 kHz  (1)________ –98 dBc/Hz ±2.44 dB
       20 kHz (2)________ –104 dBc/Hz ±2.44 dB
       30 kHz (3)________ –110 dBc/Hz ±2.44 dB
Table 3-4  Agilent E4402B Performance Verification Test Record
Agilent Technologies
Model E4402B Report No. ___________ 
Serial No. ___________ Date ___________
Test Description Minimum Results 
Measured Maximum Measurement 
Uncertainty