618 Chapter  3 
Performance Verification Test Records
Agilent E4411B Performance Verification Test Record
50Ω Termination  ___________ ___________ ___________
Minimum Loss Pad 
(Option 1DP only)
___________ ___________ ___________
1 dB Step Attenuator ___________ ___________ ___________
10 dB Step Attenuator ___________ ___________ ___________
Microwave Spectrum 
Analyzer (Option 1DN or 
1DQ only)
___________ ___________ ___________
Notes/comments: _____________________________________________________________ 
_____________________________________________________________
Table 3-16  Agilent E4411B Performance Verification Test Record
Agilent Technologies
Model E4411B Report No. ___________ 
Serial No. ___________ Date ___________
Test Description Minimum Results 
Measured Maximum Measurement 
Uncertainty
1. 10 MHz Reference 
Output Accuracy
  Settability –5.0 Hz (1)________  5.0 Hz ±293.3 µHz
3. Frequency Readout and 
Marker Frequency 
Count Accuracy
Frequency Readout 
Accuracy
 Center Freq     Span 
   1490 MHz      20 MHz 1489.83 MHz (1)________ 1490.17 MHz ±0 Hz
   1490 MHz      10 MHz 1489.91 MHz (2)________ 1490.09 MHz ±0 Hz
   1490 MHz      1 MHz 1489.991 
MHz
(3)________ 1490.009 
MHz
±0 Hz
Marker Count Accuracy
  Center Freq    Span
    1490 MHz     10 MHz 1489.999998 
MHz
(4)________ 1490.000002 
MHz
±0 Hz