Chapter  3 563
Performance Verification Test Records
Agilent E4407B Performance Verification Test Record
9. System-Related 
Sidebands
 Offset from 500 MHz 
signal 
 30 kHz to 230 kHz  (1)________ –65 dBc ±1.30 dB
 –30 kHz to –230 kHz (2)________ –65 dBc ±1.30 dB
10. Residual FM
 1 kHz Res BW,
(Non-Option 1D5)
 (1)_________ 150 Hz ±21 Hz
 1 kHz Res BW,
(Option 1D5)
 (1)_________ 100 Hz ±21 Hz
 10 Hz Res BW
(Options 1DR
and 1D5 only)
(2)_________ 2 Hz ±0.27 Hz
11. Sweep Time Accuracy
 Sweep Time
   5 ms  –1.0% (1)________ ±1.0% ±0.16%
   20 ms –1.0% (2)________ ±1.0% ±0.16%
   100 ms –1.0% (3)________ ±1.0% ±0.16%
   1 s –1.0% (4)________ ±1.0% ±0.16%
   10 s –1.0% (5)________ ±1.0% ±0.16%
   1 ms (Option AYX or B7D 
only)
–1.0% (6)________ ±1.0% ±0.16%
   500 µs (Option AYX or 
B7D only)
–1.0% (7)________ ±1.0% ±0.16%
   100 µs (Option AYX or 
B7D only)
–1.0% (8)________ ±1.0% ±0.16%
12. Display Scale Fidelity
Table 3-12  Agilent E4407B Performance Verification Test Record
Agilent Technologies
Model E4407B Report No. ___________ 
Serial No. ___________ Date ___________
Test Description Minimum Results 
Measured Maximum Measurement 