Chapter 2 411
Performance Verification Tests
59. Comms Absolute Power Accuracy (Options BAC or BAH)
1. Press Preset on the analyzer. Press the Factory Preset softkey, if it is 
displayed. Wait for the preset routine to finish. 
2. Perform a complete self-alignment and set Auto Align Off. Press 
System, Alignments, Align Now, All, and wait for the alignment routine 
to finish. Then, press Return, Auto Align, Off.
3. Zero and calibrate the power meter and power sensor connected to 
Channel A of the power meter.
4. Connect the equipment as shown in Figure 2-77.
5. Preset the synthesized signal generator. Press: 
Blue Key, Special, 0, 0 and set the signal generator as follows:
FREQUENCY, 50 MHz
AMPLITUDE, 15 dBm
6. Set the 10 dB and 1 dB step attenuators to 0 dB.
7. Obtain the actual attenuation for the 0 dB setting of each attenuator 
at 50 MHz from the metrology data for the step attenuators. In some 
cases this value might be zero, by definition. Add the two actual 
attenuations to obtain the 0 dB reference attenuation.
For example, if the actual attenuation for the 10 dB step attenuator 
at the 0 dB setting is 0.03 dB, then 10 dB Actual0dB is 0.03 dB. If the 
actual attenuation for the 1 dB step attenuator at the 0 dB setting is 
–25 –5 10 10 0 40 15
Table 2-116  GSM Transmit Power (Option BAH), Preamp Off, Input Level 
>−20 dBm
Nominal 
Input 
Level 
(dBm)
Reference 
Level 
(dBm)
Resolution 
Bandwidth 
(kHz)
Internal 
Attenuator 
(dB)
1 dB Step 
Attenuator 
(dB)
10 dB Step 
Attenuator 
(dB)
 Nominal 
Source 
Level
(dBm)
15 15 300 40 0 0 15
15 30 300 40 0 0 15
Table 2-115  cdmaOne Channel Power (Option BAC), Preamp Off, Input 
Level ≥–25 dBm
Nominal 
Input 
Level 
(dBm)
Reference 
Level 
(dBm)
Resolution 
Bandwidth 
(kHz)
Internal 
Attenuator 
(dB)
1 dB Step 
Attenuator 
(dB)
10 dB Step 
Attenuator 
(dB)
Nominal 
Source 
Level
(dBm)
RefAtten0dB 10 dB Actual0dB 1 dB Actual0dB
+=