Chapter 3 513
Performance Verification Test Records
Agilent E4404B Performance Verification Test Record
ac Coupled
Band 0
(100 kHz to 3.0 GHz)
Maximum Response (10)________ 1.0 dB ±0.11 dB
Minimum Response –1.0 dB (11)________ ±0.11 dB
Peak-to-Peak Response (12)________ 2.0 dB ±0.11 dB
ac Coupled
Band 0A
(800 MHz to 1.0 GHz)
(serial number
US39440498 or greater)
Maximum Response (13)________ 1.0 dB ±0.11 dB
Minimum Response –1.0 dB (14)________ ±0.11 dB
Peak-to-Peak Response (15)________ 2.0 dB ±0.11 dB
ac Coupled
Band 0B
(1.7 GHz to 2.0 GHz)
(serial number
US39440498 or greater)
Maximum Response (16)________ 1.0 dB ±0.11 dB
Minimum Response –1.0 dB (17)________ ±0.11 dB
Peak-to-Peak Response (18)________ 2.0 dB ±0.11 dB
ac Coupled
Band 1
(3.0 GHz to 6.7 GHz)
Maximum Response (19)________ 2.5 dB ±0.39 dB
Minimum Response –2.5 dB (20)________ ±0.39 dB
Table 3-8 Agilent E4404B Performance Verification Test Record
Agilent Technologies
Model E4404B Report No. ___________
Serial No. ___________ Date ___________
Test Description Minimum Results
Measured Maximum Measurement
Uncertainty