354 Chapter 2
Performance Verification Tests
44. Tracking Generator Level Flatness: Agilent E4402B, E4403B, E4404B, E4405B, E4407B,
and E4408B (Option 1DN)
2. Set the analyzer resolution bandwidth to 10 kHz by pressing
BW/Avg, Res BW, 10 kHz.
Repeat step 3 through step 7 for each Center Frequency value in
Table 2-98.
3. Set the analyzer center frequency to 9 kHz, by pressing FREQUENCY,
9 kHz.
4. Press Single.
5. Record the DVM readout in Table 2-98.
6. Subtract the 100 kHz Level Flatness readout in Table 2-97 from the
100 kHz DVM Readout in Table 2-98 and record as the DVM Offset
at 100 kHz.
DVM Offset at 100 kHz ___________ dB
For example, if the Level Flatness reading from Table 2-97 is
0.7 dB and the DVM Readout from Table 2-98 is 0.53 dBm, the
DVM offset would be 1.23 dB.
7. Add the DVM Offset at 100 kHz from step 6, above, to each of the
DVM Readouts in Table 2-98 and record as the Corrected Level
Flatness in Column 3.
For example, if the DVM Readout from Table 2-98 is 0.22 dBm,
and the DVM Offset is 1.23 dB, the Corrected Level Flatness
would be 1.01 dB.
8. Press System, Alignments, Auto Align, All.
Set to 50 impedance SMATH 10a
Set to dBm MATH 5a
Set to Synchronous SETACV 3a
Sub-sampled mode
a. To set the Agilent 3458A multimeter functions from the
front panel, press the blue shift key, then Recall State (T)
key. Use the (up) and (down) arrows to select the
appropriate function, then enter the value from the
numeric keypad and press enter.
Parameter Setting

DVM Offset DVM Readout= Level Flatness

Corrected Level Flatness DVM Readout= DVM Offset+