568 Chapter 3
Performance Verification Test Records
Agilent E4407B Performance Verification Test Record
–4 –2.0% (93)________ 2.0% ±0.064%
–8 –2.0% (94)________ 2.0% ±0.064%
–12 –2.0% (95)________ 2.0% ±0.064%
–16 –2.0% (96)________ 2.0% ±0.064%
–20 –2.0% (97)________ 2.0% ±0.063%
Linear Fidelity,
Res BW 300 Hz
(Option 1DR only)
dB from Ref Level
–4 –2.0% (98)________ 2.0% ±0.064%
–8 –2.0% (99)________ 2.0% ±0.064%
–12 –2.0% (100)________ 2.0% ±0.064%
–16 –2.0% (101)________ 2.0% ±0.064%
–20 –2.0% (102)________ 2.0% ±0.063%
Zero Span,
Res BW 300 Hz
(Option 1DR only)
dB from Ref Level
–4 –0.36 dB (103)_________ 0.36 dB ±0.082 dB
–8 –0.42 dB (104)________ 0.42 dB ±0.078 dB
–12 –0.48 dB (105)________ 0.48 dB ±0.075 dB
–16 –0.54 dB (106)________ 0.54 dB ±0.073 dB
–20 –0.60 dB (107)________ 0.60 dB ±0.078 dB
–24 –0.66 dB (108)________ 0.66 dB ±0.074 dB
–28 –0.72 dB (109)________ 0.72 dB ±0.073 dB
–32 –0.78 dB (110)________ 0.78 dB ±0.077 dB
–36 –0.84 dB (111)________ 0.84 dB ±0.075 dB
Table 3-12 Agilent E4407B Performance Verification Test Record
Agilent Technologies
Model E4407B Report No. ___________
Serial No. ___________ Date ___________
Test Description Minimum Results
Measured Maximum Measurement
Uncertainty