624 Chapter 3
Performance Verification Test Records
Agilent E4411B Performance Verification Test Record
–72 –0.4 dB (85)________ 0.4 dB ±0.084 dB
–76 –0.4 dB (86)________ 0.4 dB ±0.083 dB
–80 –0.4 dB (87)________ 0.4 dB ±0.089 dB
Linear Fidelity,
Res BW 1 kHz
dB from Ref Level
–4 –2.0% (93)________ 2.0% ±0.064%
–8 –2.0% (94)________ 2.0% ±0.064%
–12 –2.0% (95)________ 2.0% ±0.064%
–16 –2.0% (96)________ 2.0% ±0.064%
–20 –2.0% (97)________ 2.0% ±0.063%
Linear Fidelity,
Res BW 300 Hz
(Option 1DR only)
dB from Ref Level
–4 –2.0% (98)________ 2.0% ±0.064%
–8 –2.0% (99)________ 2.0% ±0.064%
–12 –2.0% (100)________ 2.0% ±0.064%
–16 –2.0% (101)________ 2.0% ±0.064%
–20 –2.0% (102)________ 2.0% ±0.063%
Zero Span,
Res BW 300 Hz
(Option 1DR only)
dB from Ref Level
–4 –0.36 dB (103)_________ 0.36 dB ±0.082 dB
–8 0.42 dB (104)________ 0.42 dB ±0.078 dB
–12 –0.48 dB (105)________ 0.48 dB ±0.075 dB
–16 –0.54 dB (106)________ 0.54 dB ±0.073 dB
Table 3-16 Agilent E4411B Performance Verification Test Record
Agilent Technologies
Model E4411B Report No. ___________
Serial No. ___________ Date ___________
Test Description Minimum Results
Measured Maximum Measurement
Uncertainty