Chapter 3 531
Performance Verification Test Records
Agilent E4405B Performance Verification Test Record
Cumulative Log Fidelity,
Res BW 1 kHz
dB from Ref Level
–4 –0.30 dB (1)________ 0.30 dB ±0.082 dB
–8 –0.30 dB (2)________ 0.30 dB ±0.078 dB
–12 –0.40 dB (3)________ 0.40 dB ±0.075 dB
–16 –0.40 dB (4)________ 0.40 dB ±0.073 dB
–20 –0.40 dB (5)________ 0.40 dB ±0.078 dB
–24 –0.50 dB (6)________ 0.50 dB ±0.074 dB
–28 –0.50 dB (7)________ 0.50 dB ±0.073 dB
–32 –0.60 dB (8)________ 0.60 dB ±0.077 dB
–36 –0.60 dB (9)________ 0.60 dB ±0.075 dB
–40 –0.60 dB (10)________ 0.60 dB ±0.081 dB
–44 –0.70 dB (11)________ 0.70 dB ±0.077 dB
–48 –0.70 dB (12)________ 0.70 dB ±0.076 dB
–52 –0.70dB (13)________ 0.70 dB ±0.080 dB
–56 –0.70 dB (14)________ 0.70 dB ±0.078 dB
–60 –0.70 dB (15)________ 0.70 dB ±0.084 dB
–64 –0.80 dB (16)________ 0.80 dB ±0.081 dB
–68 –0.80 dB (17)________ 0.80 dB ±0.080 dB
–72 –0.80 dB (18)________ 0.80 dB ±0.084 dB
–76 –0.80 dB (19)________ 0.80 dB ±0.083 dB
–80 –0.80 dB (20)________ 0.80 dB ±0.089 dB
–84 –1.15 dB (21)________ 1.15 dB ±0.086 dB
Table 3-10 Agilent E4405B Performance Verification Test Record
Agilent Technologies
Model E4405B Report No. ___________
Serial No. ___________ Date ___________
Test Description Minimum Results
Measured Maximum Measurement
Uncertainty