Chapter 2 355
Performance Verification Tests
44. Tracking Generator Level Flatness: Agilent E4402B, E4403B, E4404B, E4405B, E4407B,
and E4408B (Option 1DN)
1. Locate the most positive Level Flatness reading in Table 2-97 and
Table 2-98 for frequencies <1 MHz and enter this value as Test
Record entry 1 of the performance verification test record.
2. Locate the most negative Level Flatness reading in Table 2-97 and
Table 2-98 for frequencies <1 MHz and enter this value as Test
Record entry 2 of the performance verification test record.
3. Locate the most positive Level Flatness reading in Table 2-97 and
Table 2-98 for frequencies 1 MHz and 10 MHz and enter this value
as Test Record entry 3 of the performance verification test record.
4. Locate the most negative Level Flatness reading in Table 2-97 and
Table 2-98 for frequencies 1 MHz and 10 MHz and enter this value
as Test Record entry 4 of the performance verification test record.
5. Locate the most positive Level Flatness reading in Table 2-97 for
frequencies 10 MHz and 1.5 GHz and enter this value as Test
Record entry 5 of the performance verification test record.
6. Locate the most negative Level Flatness reading in Table 2-97 for
frequencies 10 MHz and 1.5 GHz and enter this value as Test
Record entry 6 of the performance verification test record.
7. Locate the most positive Level Flatness reading in Table 2-97 for
frequencies >1.5 GHz and enter this value as Test Record entry 7 of
the performance verification test record.
8. Locate the most negative Level Flatness reading in Table 2-97 for
frequencies >1.5 GHz and enter this value as Test Record entry 8 of
the performance verification test record.
Table 2-98 Tracking Generator Level Flatness Worksheet, 100 kHz
Center
Frequency
DVM Readout
(dBm)
Corrected Level
Flatness
(dB)
9 kHz
20 kHz
40 kHz
60 kHz
80 kHz
100 kHz