44 Chapter 2
Performance Verification Tests
1. 10 MHz Reference Output Accuracy
the maximum frequency change.
16.Divide the maximum frequency change by two and record the result
as the settability.
Table 2-3 10 MHz Reference Accuracy Worksheet
Description Measurement
Counter Reading 1 ___________ Hz
Timebase Fine DAC ___________ Hz
Counter Reading 2 ___________ Hz
Counter Reading 3 ___________ Hz
Positive Frequency Change ___________ Hz
Negative Frequency Change ___________ Hz
Maximum Frequency Change ___________ Hz
Settability ___________ Hz