518 Chapter  3 
Performance Verification Test Records
Agilent E4404B Performance Verification Test Record
10 Hz RBW, Preamp On,
20 to 30° C
          10 MHz to 1 GHz (20)________ –151 dBm ±1.23 dB
          1 GHz to 2 GHz (21)________ –151 dBm ±1.23 dB
          2 GHz to 3 GHz (22)________ –149 dBm ±1.23 dB
39. Residual Responses
150 kHz to 6.7 GHz (1)________ –90 dBm ±0.87 dB
40. Fast Time Domain 
Amplitude Accuracy 
(Option AYX only)
  Amplitude Error –0.3% (1)________ 0.3% ±0.029%
42. Tracking Generator 
Absolute Amplitude 
and Vernier Accuracy 
Option 1DN only)
Absolute Amplitude 
Accuracy
–0.75 dB (1)________ 0.75 dB ±0.096 dB
Absolute Vernier 
Accuracy
–2 dB –0.5 dB (2)________ 0.5 dB ±0.008 dB
–3 dB –0.5 dB (3)________ 0.5 dB ±0.008 dB
–5 dB –0.5 dB (4)________ 0.5 dB ±0.008 dB
–6 dB –0.5 dB (5)________ 0.5 dB ±0.008 dB
–7 dB –0.5 dB (6)________ 0.5 dB ±0.008 dB
–8 dB –0.5 dB (7)________ 0.5 dB ±0.008 dB
–9 dB –0.5 dB (8)________ 0.5 dB ±0.008 dB
–10 dB –0.5 dB (9)________ 0.5 dB ±0.008 dB
Incremental Vernier 
Accuracy
Table 3-8  Agilent E4404B Performance Verification Test Record
Agilent Technologies
Model E4404B Report No. ___________ 
Serial No. ___________ Date ___________
Test Description Minimum Results 
Measured Maximum Measurement 
Uncertainty