Chapter 2 377
Performance Verification Tests
49. Tracking Generator LO Feedthrough: Agilent E4402B, E4403B, E4404B, E4405B, E4407B,
and E4408B(Option 1DN)
five minutes before proceeding.
Complete step 2 only if more than 24 hours have elapsed since
performing a front-panel calibration of the microwave analyzer.
2. Perform a front-panel calibration of the microwave analyzer by
performing the following steps:
a. Connect a BNC cable between CAL OUTPUT and 50 Input.
b. Press CAL, REALIGN LO & IF.
c. Set FREQUENCY, 300 MHz.
d. Set SPAN, 20 MHz.
e. Set AMPLITUDE, –10 dBm.
f. Press PEAK SEARCH.
g. Press CAL, REF LVL ADJ and use the ⇑ ⇓ arrows to adjust the DAC
value to a marker amplitude reading of 10 dBm. Press
STORE REF LVL.
h. Disconnect the BNC cable from between the CAL OUTPUT and
50 Input.
3. Press Preset on the analyzer under test. Press the Factory Preset
softkey, if it is displayed.
4. Use the type-N cable to connect the RF Input to the tracking
generator RF OUT on the analyzer under test as shown in Figure
2-69. Do not connect to the microwave analyzer RF Input yet.
5. Initialize the test equipment by pressing the following keys on the
analyzer under test:
FREQUENCY, 50 MHz
SPAN, Zero Span
BW / Avg, 30 kHz
System, Alignments, Auto Align, Off
Marker
Source, Amplitude On, –5 dBm
Source, Tracking Peak (Wait for the PEAKING SIGNAL message to
disappear.)
FREQUENCY, 9 kHz
Source, Amplitude On, –2 dBm
Single
6. On the microwave analyzer, press the following keys:
FREQUENCY, 3.921409 GHz
SPAN, 100 kHz
AMPLITUDE, 0 dBm
BW, 1 kHz
7. Disconnect the type-N cable from between the analyzer RF INPUT