Performance Verification Test Records
Agilent E4403B Performance Verification Test Record
23. Frequency Response  Note: Enter data in the appropriate section below depending upon 
the ambient temperature at which the test was performed. 
 20 to 30° C
   Maximum Response (1)________ 0.5 dB ±0.19 dB
   Minimum Response –0.5 dB (2)________ ±0.19 dB
   Peak-to-Peak Response (3)________ 1.0 dB ±0.19 dB
 0 to 55° C
   Maximum Response (1)________ 1.0 dB ±0.19 dB
   Minimum Response –1.0 dB (2)________ ±0.19 dB
   Peak-to-Peak Response (3)________ 2.0 dB ±0.19 dB
29. Other Input-Related 
Spurious Responses
   Input Frequency
        2042.8 MHz (1)________ –65 dBc ±0.891 dB
        2642.8 MHz (2)________ –65 dBc  ±0.891 dB
        1820.8 MHz (3)________ –65 dBc ±0.891 dB
        278.5 MHz (4)________ –65 dBc ±0.891 dB
31. Spurious Responses  Note: Entry 2 does not apply to the Agilent E4403B.
   300 MHz TOI 7.5 dBm (1)________   ±0.93 dB
   300 MHz SHI 30 dBm (3)________ ±0.41 dB
   900 MHz SHI 40 dBm (4)________ ±0.41 dB
33. Gain Compression Note: Entry 2 does not apply to the Agilent E4403B.
Test Frequency
  53 MHz (1)________  1.0 dB ±0.123 dB
  1403 MHz (3)________  1.0 dB ±0.123 dB
  2503 MHz (4)________  1.0 dB ±0.123 dB
Table 3-6  Agilent E4403B Performance Verification Test Record
Agilent Technologies
Model E4403B Report No. ___________ 
Serial No. ___________ Date ___________
Test Description Minimum Results 
Measured Maximum Measurement 
Uncertainty