
Chapter  3 549
Performance Verification Test Records
Agilent E4405B Performance Verification Test Record
 1403 MHz (3)________  1.0 dB ±0.123 dB
 2503 MHz (4)________  1.0 dB ±0.123 dB
 4403 MHz (5)________  1.0 dB ±0.123 dB
 7603 MHz (6)________  1.0 dB ±0.123 dB
37. Displayed Average 
Noise Level
Note: Enter results with preamp on in the appropriate section based 
upon the ambient temperature when the test was performed.
1 kHz RBW, Preamp Off
          10 MHz to 1 GHz (1)________ –116 dBm ±1.23 dB
          1 GHz to 2 GHz (2)________ –116 dBm ±1.23 dB
          2 GHz to 3 GHz (3)________ –112 dBm ±1.23 dB
          3 GHz to 6 GHz (4)________ –112 dBm ±1.23 dB
          6 GHz to 12 GHz (5)________ –111 dBm ±1.23 dB
       12 GHz to 13.2 GHz (6)________ –107 dBm ±1.23 dB
1 kHz RBW, Preamp On,
0 to 55° C
          10 MHz to 1 GHz (7)________ –131 dBm ±1.23 dB
          1 GHz to 2 GHz (8)________ –131 dBm ±1.23 dB
          2 GHz to 3 GHz (9)________ –127 dBm ±1.23 dB
10 Hz RBW, Preamp Off
          10 MHz to 1 GHz (10)________ –135 dBm ±1.23 dB
          1 GHz to 2 GHz (11)________ –135 dBm ±1.23 dB
          2 GHz to 3 GHz (12)________ –131 dBm ±1.23 dB
          3 GHz to 6 GHz (13)________ –131 dBm ±1.23 dB
          6 GHz to 12 GHz (14)________ –130 dBm ±1.23 dB
       12 GHz to 13.2 GHz (15)________ ��120 dBm ±1.23 dB
Table 3-10  Agilent E4405B Performance Verification Test Record
Agilent Technologies
Model E4405B Report No. ___________ 
Serial No. ___________ Date ___________
Test Description Minimum Results 
Measured Maximum Measurement 
Uncertainty