Calibration Guide
Agilent Technologies ESA Spectrum Analyzers
Notice
Page
Page
Page
Page
Page
Page
Page
Page
Calibration Cycle
Performance Verification Test Tables
Table 1-1 Agilent E4401B Performance Verification Tests
Page
Table 1-2 Agilent E4402B Performance Verification Tests
Table 1-3 Agilent E4403B Performance Verification Tests
Page
Table 1-4 Agilent E4404B Performance Verification Tests
Page
Table 1-5 Agilent E4405B Performance Verification Tests
Page
Table 1-6 Agilent E4407B Performance Verification Tests
Table 1-7 Agilent E4408B Performance Verification Tests
Table 1-8 Agilent E4411B Performance Verification Tests
Before You Start
Recording the test results
Performing Self-Alignment
Periodic verification of operation
Test equipment you will need
Page
< < < <
Page
Page
Page
Table 1-11 Recommended Adapters
Table 1-12 Recommended Cables
Page
Tests included in this section:
Page
Page
Calibration
Page
Table 2-1 Calibration Requirements
Page
Table 2-2 Agilent 11713A Settings for Agilent 8494G and Agilent 8496G
1. 10 MHz Reference Output Accuracy
Page
Page
Page
Page
5 Minute Warm-up Error Counter Reading 1 Counter Reading 3()10=
15 Minute Warm-up Error Counter Reading 2 Counter Reading 3()10=
Page
Page
Page
Page
Page
Page
Page
5. Frequency Span Accuracy: Agilent E4401B and E4411B
100 kHz and 100 MHz Frequency Span Readout Accuracy
Page
Table 2-8 Frequency Span Readout Accuracy
6. Frequency Span Accuracy: Agilent E4402B, E4403B, E4404B, E4405B,
Page
Page
Table 2-9 Frequency Span Readout Accuracy
7. Noise Sidebands
Page
Page
Page
Page
Page
Page
8. Noise Sidebands - Wide Offsets: Agilent E4402B, E4404B, E4405B and E4407B
Page
Page
Page
Page
9. System-Related Sidebands
Equipment Required
Additional Equipment for 75 Input
Procedure
Page
Page
10. Residual FM
Procedure
Part 1: Residual FM
Page
Part 2: Residual FM for Option 1DR and 1D5
Page
11. Sweep Time Accuracy
Page
Sweep Time Accuracy 100 Mkr1 0.8 Sweep Time() Sweep Time
------------------------------------------------------------------------
12. Display Scale Fidelity
Page
Part 1: Log Display Scale Fidelity, Analog Bandwidths
CLFE Total Actual Attenuation Mkr Reading Total Actual Atten 0 dB from Ref Level()+=
90 Chapter 2
Column 8
Column 9
Test Record Entry CLFE (dB)
Test Record Entry ILFE (dB)
Part 2: Log Display Scale Fidelity, Digital Bandwidths
92 Chapter 2
BW/Avg, Average (On) 5, Enter
CLFE Total Actual Attenuation MkrReading Total Actual Atten
0 dB from R
(+=
Chapter 2 93
Part 3: Linear Display Scale Fidelity, Analog Bandwidths
System, Alignments, Auto Align, Off FREQUENCY, 50 MHz
ILFE CLFE current()CLFE previous()=
Ideal Mkr Reading mV()1000 0.05 10ATmeasATref+()10
Ideal Mkr Reading(mV) 1000 0.05 107.9820.012+()1089.3 mV==
Part 4: Linear Display Scale Fidelity, Digital Bandwidths
This section is for analyzers with Option 1DR (narrow resolution bandwidths) only.
LFE % of RL()100 Actual Mkr Reading Ideal Mkr Reading 223.6 mV
----------------------------------------------------------------------------------------------------
Actual Mkr Reading 85.0 mV= Ideal Mkr Reading 89.3 mV= LFE %of RL()100 85.0 89.3 223.6
Ideal Mkr Reading mV()1000 0.05 10ATmeasATref+()10
LFE %of RL()100 Actual Mkr Reading Ideal Mkr Reading 223.6 mV
----------------------------------------------------------------------------------------------------
Chapter 2 97
10.If the analyzer has a 75W Input press:
Input/Output (or Input) Input Z Corr (75)
15. On the analyzer press Peak Search (or Search), Marker >, and Marker > CF.
Table 2-16 Linear Scale Fidelity Worksheet, Option 1DR Narrow Resolution Bandwidths
Page
Table 2-17 Zero Span Cumulative Log Fidelity Worksheet, Opt 1DR Narrow Resolution Bandwidths
13. Input Attenuation Switching Uncertainty
Page
Table 2-18 Actual Attenuation Worksheet
A
7. Preset the synt hesized signal generato r ( set the controls as follows:
Page
Page
14. Reference Level Accuracy: Agilent E4401B and E4411B
Actual Attenuator Step X dB()Actual Attenuation 20 dB()Actual Attenuation X dB()()=
Page
Page
Page
Page
Page
Page
Page
Page
15. Reference Level Accuracy: Agilent E4402B, E4403B, E4404B, E4405B, E4407B, and E4408B.
Actual Attenuator Step X dB()Actual Attenuation 20 dB()Actual Attenuation X dB()()=
Page
performance test record. The following is an example for 30 dBm reference level:
Log Scale, Digital Bandwidths, Option 1DR
8. Set the 1 dB step attenuator to set the signal peak 1 to 3 divisions below the reference level.
Table 2-24 Log Mode, Analog Bandwidths Worksheet
Page
Page
Page
Table 2-27 Linear Mode, Digital Bandwidths Worksheet, Option 1DR
16. Resolution Bandwidth Switching Uncertainty
Equipment Required for Agilent E4402B, E4403B, E4404B, E4405B, E4407B, and E4408B
Page
Page
Page
17. Absolute Amplitude Accuracy (Reference Settings): Agilent E4401B and E4411B
Page
Page
Page
Page
Page
Page
18. Absolute Amplitude Accuracy (Reference Settings): Agilent E4402B, E4403B, E4404B,
Page
Page
Page
Page
Page
Page
19. Overall Absolute Amplitude Accuracy:
Page
Page
Amplitude Accuracy Measured Amplitude Amptd
Page
Page
Page
20. Overall Absolute Amplitude Accuracy: Agilent E4402B, E4403B, E4404B, E4405B,
Page
Page
Amplitude Accuracy Measured Amplitude Amptd
Page
Page
Page
21. Resolution Bandwidth Accuracy
Page
Page
Page
6 dB Bandwidth Upper Marker Frequency Lower Marker Frequency=
22. Frequency Response: Agilent E4401B and E4411B
Page
Page
Table 2-40 Source/Splitter Characterization
a. These values do not apply to analyzers with 75 inputs (Option 1DP).
Page
Page
a. These values do not apply to analyzers with 75 inputs (Option 1DP).
Page
Page
Page
23. Frequency Response, Agilent E4402B and E4403B
Page
Page
Table 2-44 Source/Splitter Characterization
Page
Page
Page
Page
Table 2-45 Frequency Response Worksheet, 100 kHz to 3.0 GHz
Table 2-45 Frequency Response Worksheet, 100 kHz to 3.0 GHz
Page
Page
Page
Page
Page
24. Frequency Response, Agilent E4404B,
Page
Page
Table 2-49 Source/Splitter Characterization
Page
Page
Page
Page
Table 2-50 Frequency Response Worksheet, 100 kHz to 3.0 GHz
Table 2-50 Frequency Response Worksheet, 100 kHz to 3.0 GHz
Page
Page
a. These frequencies apply only to analyzers with Option UKB.
Table 2-51 Frequency Response Worksheet, 100 kHz
Page
Page
Page
Page
Table 2-52 Frequency Response Worksheet, >3 GHz
Page
Page
Page
Page
Page
Page
Page
Page
25. Frequency Response (Preamp On): Agilent E4401B
Page
Page
Page
Page
Page
Page
Page
26. Frequency Response (Preamp On): Agilent E4402B
Page
Page
Page
Page
Page
Page
Table 2-61 Frequency Response Worksheet, Preamp On
Page
Page
27. Frequency Response (Preamp On): Agilent E4404B, E4405B, and E4407B
Page
Page
a. This entry is for data taken with the function generator as source.
Table 2-63 Source/Splitter Characterization
Page
Page
Page
Page
Table 2-64 Frequency Response Worksheet, Preamp On
Page
Page
Page
Page
28. Other Input-Related Spurious Responses:
Page
Page
Page
Table 2-67 Other Input-Related Spurious Responses Worksheet
a. Image response b. Multiple response
Page
Page
Page
Page
30.On the analyzer, press AMPLITUDE, Presel Center. 31.On the analyzer, press the following:
a. Image response b. Multiple response c. Out-of-band response
Table 2-68 Other Input-Related Spurious Responses
30. Spurious Responses: Agilent E4401B and E4411B
Page
Page
Page
Page
Page
Page
Page
31. Spurious Responses: Agilent E4402B and E4403B
Page
Page
Page
Page
Page
0.60 dB 7.05 dBm 6.45 dBm()=
0.70 dB 7.05 dBm 6.35 dBm()=
38.4 dBm 20 dBm 59 dB() 0.60 dB()+=
52.3 dBm 20 dBm 73 dB() 0.70 dB()+=
32. Spurious Responses: Agilent E4404B,
Page
Page
Page
Page
Page
Page
Page
0.60 dB 7.05 dBm 6.45 dBm()=
0.70 dB 7.05 dBm 6.35 dBm()=
1.2 dB 7.35 dBm 6.15 dBm()=
1.2 dB 7.25 dBm 6.05 dBm()=
Page
Page
Page
Page
Page
33. Gain Compression: Agilent E4401B, E4402B, E4403B, and E4411B
Page
Table 2-77 Source Frequency and Amplitude Settings
a. Agilent E4401B and E4402B with Option 1DR only. b. Agilent E4402B and E4403B only.
Page
34. Gain Compression: Agilent E4404B,
Page
Table 2-80 for each frequency indicated. Table 2-79 Source Frequency and Amplitude Settings
Page
Page
35. Displayed Average Noise Level:
Page
Ref Lvl Offset (1 kHz RBW) Ref Amptd Meas Amptd (1 kHz RBW)=
Page
Page
Table 2-81 DANL Measurement Sequence, E4401B
Table 2-82 DANL Measurement Sequence, E4411B
Table 2-83 DANL Measurement Sequence, E4401B 75 Inputs
Table 2-82 DANL Measurement Sequence, E4411B
Page
Page
Page
36. Displayed Average Noise Level: Agilent E4402B and E4403B
Page
Ref Lvl Offset(1 kHz RBW) Ref Amptd Meas Amptd(1 kHz RBW)=
Page
Subset D in Table 2-85.
8. After performing all applicable DANL measurement procedures,
Table 2-85 DANL Measurement Sequence, E4402B
Page
Page
37. Displayed Average Noise Level: Agilent E4404B and E4405B
Page
Ref Lvl Offset (1 kHz RBW) Ref Amptd Meas Amptd (1 kHz RBW)=
Ref Lvl Offset (10 Hz RBW) Ref Amptd Meas Amptd (10 Hz RBW)=
Page
Table 2-87 DANL Measurement Sequence, E4404B
Page
Page
38. Displayed Average Noise Level: Agilent E4407B and E4408B
Page
Ref Lvl Offset(1 kHz RBW) Ref Amptd Meas Amptd(1 kHz RBW)=
Page
Subset D in Table 2-89. Table 2-89 DANL Measurement Sequence, E4407B
Page
Page
39. Residual Responses
Page
Page
Page
Amplitude Error Mkr1 1.0()100= Amplitude Error Mkr1 100= %
Equipment Required
Additional Equipment for Option 1DQ
Page
Vernier Accuracy Measured Power Level (dB) Source Vernier Setting (dB)=
Power Sweep Accuracy Positive Power Sweep Accuracy Negative Power Sweep Accuracy=
Table 2-92 Vernier Accuracy Worksheet
Page
Page
Page
Vernier Accuracy Measured Power Level (dB) Source Vernier Setting (dB)( 4 dB )+=
43. Tracking Generator Level Flatness: Agilent E4401B and E4411B
dBm 10 log10 E2R()1mW()=
Additional Equipment for Option 1DQ
Procedure Tracking Generator Level Flatness, Center Frequency 100 kHz
Page
Page
Tracking Generator Level Flatness, Center Frequency 100 kHz
DVM Offset DVM Readout= Level Flatness
Corrected Level Flatness DVM= DVM Offset+
Page
dBm 10 log10 E2R()1mW()=
Tracking Generator Level Flatness, Center Frequency 100 kHz
Page
Tracking Generator Level Flatness, Center Frequency 100 kHz
DVM Offset DVM Readout= Level Flatness
Corrected Level Flatness DVM Readout= DVM Offset+
Page
45. Tracking Generator Harmonic Spurious Outputs: Agilent E4401B and E4411B
Equipment Required
Additional Equipment Required for Option 1DQ
Procedure
Page
Page
Page
Page
Page
Page
Page
47. Tracking Generator Non-Harmonic Spurious Outputs: Agilent E4401B and E4411B
Page
Measuring Fundamental Amplitudes
Measuring Non-Harmonic Responses
Page
Non-harmonic Amplitude Marker Amplitude Fundamental Amplitude=
Determining the Highest Non-harmonic Spurious Response
Page
Page
Page
Non-harmonic Amplitude Marker Amplitude Fundamental Amplitude=
Page
Page
Page
Page
Page
Page
Figure 2-70 Gate Delay and Gate Length Test Setup
3. Set up the function generator as follows:
4. Press
Page
Page
Table 2-106 Gate Delay and Gate Length Accuracy Worksheet
Additional Equipment for Option 1DP
Page
Page
52. First LO OUTPUT Power Accuracy (Option AYZ only)
Page
Table 2-107 First LO Output Amplitude Accuracy Worksheet
53. IF INPUT Accuracy (Option AYZ only)
Page
New CW Fre
q
uenc
y
321.4 MHz Mkr1 Fre
Page
54. Comms Frequency Response (Option BAC or BAH)
Page
Page
Chapter 2 397
54. Comms Frequency Response (Option BAC or BAH)
Table 2-108 Source/Splitter Characterization
398 Chapter 2
54. Comms Frequency Response (Option BAC or BAH)
Page
Page
Flatness Error Mkr1 Amptd Current Channel B Splitter Tracking Error=
Page
Table 2-110 Frequency Response Worksheet Test:_______ Atten:_______ RBW:_______ Preamp:_______
Page
Page
58. GSM Phase and Frequency Error (Options BAH and B7E)
Page
Page
59. Comms Absolute Power Accuracy (Options BAC or BAH)
Page
Page
Page
Amplitude Accuracy Measured Amplitude Amptd
A
mplitude Accuracy Measured Amplitude Amptd15dBm ActualTotalAtten RefAtten0dB
+()=
Measured Amplitude Amptd + ()
<
Table 2-120 GSM Transmit Power (Option BAH), Preamp Off, Input level
Table 2-121 Amplitude Accuracy Worksheet, cdmaOne, Input Level <25 dBm
Table 2-122 Amplitude Accuracy Worksheet, GSM, Input Level 20 dBm
Page
Table 2-124 cdmaOne Channel Power Accuracy (PCS Band)
Table 2-125 GSM Transmit Power Accuracy (Cellular Band)
Table 2-126 GSM Transmit Power Accuracy (PCS Band)
RefAtten40dB 10 dB Actual40dB 1 dB Actual0dB
Page
Measured Amplitude Amptd
+() 40dB+
Table 2-129 cdmaOne Receive Channel Power (Cellular Band) Preamp Off
Table 2-130 cdmaOne Receive Channel Power (PCS Band) Preamp Off
Page
RefAtten40dB 10 dB Actual40dB 1 dB Actual0dB
Page
Measured Amplitude Amptd
+() 40dB+
Table 2-134 cdmaOne Receive Channel Power (PCS Band) Preamp On
Table 2-133 cdmaOne Receive Channel Power (Cellular Band) Preamp On
Page
Agilent E4401B Performance Verification Test
Tests for the Agilent E4401B only are included in this test record,
Table 3-1 Agilent E4401B Performance Verification Test Record
Page
Page
Page
Page
Page
Page
Page
Page
Page
Page
Table 3-2 Agilent E4401B Performance Verification Test Record
Page
Page
Page
Page
Page
Page
Page
Page
Page
Agilent E4402B Performance Verification Test
Tests for the Agilent E4402B only are included in this test record,
Table 3-3 Agilent E4402B Performance Verification Test Record
Page
Page
Page
Page
Page
Page
Page
Page
Page
Page
Page
Page
Page
Page
Page
Page
Page
Page
Page
Page
Page
Page
Page
Page
Page
30 40 60
70 80
Page
Agilent E4403B Performance Verification
Tests for the Agilent E4403B only are included in this test record,
Table 3-5 Agilent E4403B Performance Verification Test Record
Page
Page
Page
Page
Page
Page
Page
Page
Page
Page
Page
Page
Page
Page
Agilent E4404B Performance Verification Test
Tests for the Agilent E4404B only are included in this test record,
Table 3-7 Agilent E4404B Performance Verification Test Record
Page
Page
Page
Page
Page
Page
Page
Page
Page
Page
Page
Page
Page
Page
Page
Page
Page
Page
Page
Page
Page
Page
Page
Page
Page
Page
Page
Page
Agilent E4405B Performance Verification
Tests for the Agilent E4405B only are included in this test record,
Table 3-9 Agilent E4405B Performance Verification Test Record
Page
Page
Page
Page
Page
Page
Page
Page
Page
Page
Page
Page
Page
Page
Page
Page
Page
Page
Page
Page
Page
Page
Page
Page
Page
Page
Page
Page
Page
Page
Agilent E4407B Performance Verification
Tests for the Agilent E4407B only are included in this test record,
Table 3-11 E4407B Performance Verification Test Record
Page
Page
Page
Page
Page
Page
Page
Page
Page
Page
Page
Page
Page
Page
Page
Page
Page
Page
Page
Page
Page
Page
Page
Page
Page
Page
Page
Page
Page
Page
Page
Page
Page
Page
Page
Page
Page
Agilent E4408B Performance Verification Test
Tests for the Agilent E4408B only are included in this test record,
Table 3-13 Agilent E4408B Performance Verification Test Record
Page
Page
Page
Page
Page
Page
Page
Page
Page
Page
Page
Page
Page
Page
Page
Page
Page
Page
Page
Agilent E4411B Performance Verification Test
Tests for the Agilent E4411B only are included in this test record,
Table 3-15 Agilent E4411B Performance Verification Test Record
Page
Page
Page
Page
Page
Page
Page
Page
Page
Page
Page
Page
Page
Page
Page
Page
Page
What Youll Find in This Chapter
Before You Call Agilent Technologies
Check the Basics
Read the Warranty
Service Options
How to Call Agilent Technologies
Chapter 4 639
If You Have a Problem Before You Call Agilent Technologies
Table 4-1 Agilent Technologies Sales and Service Offices
How to Return Your Analyzer for Service
Service Tag
Original Packaging
Page
Other Packaging