Chapter 2 51
Performance Verification Tests
4. Frequency Readout and Marker Frequency Count Accuracy: Agilent E4404B, E4405B,
E4407B, and E4408B
4. Frequency Readout and Marker Frequency
Count Accuracy: Agilent E4404B, E4405B,
E4407B, and E4408B
The frequency readout accuracy of the analyzer is tested with an input 
signal of known frequency. By using the same frequency standard for 
the analyzer and the synthesized sweeper, the frequency reference error 
is eliminated.
There are no related adjustment procedures for this performance test. 
Equipment Required 
Synthesized sweeper
Adapter, Type N (m) to APC 3.5 (f)
Adapter, APC 3.5 (f) to APC 3.5 (f)
Cable, APC 3.5, 91 cm (36 in)
Cable, BNC, 122 cm (48 in)
Additional Equipment for Option BAB 
Adapter, APC 3.5 (f) to APC 3.5 (f)
Figure 2-4  Frequency Readout and Marker Count Accuracy Test Setup