Chapter 3 435
Performance Verification Test Records
Agilent E4401B Performance Verification Test Record
–12 –0.4 dB (70)________ 0.4 dB ±0.075 dB
–16 –0.4 dB (71)________ 0.4 dB ±0.073 dB
–20 –0.4 dB (72)________ 0.4 dB ±0.078 dB
–24 –0.4 dB (73)________ 0.4 dB ±0.074 dB
–28 –0.4 dB (74)________ 0.4 dB ±0.073 dB
–32 –0.4 dB (75)________ 0.4 dB ±0.077 dB
–36 –0.4 dB (76)________ 0.4 dB ±0.075 dB
–40 –0.4 dB (77)________ 0.4 dB ±0.081 dB
–44 –0.4 dB (78)________ 0.4 dB ±0.077 dB
–48 –0.4 dB (79)________ 0.4 dB ±0.076 dB
–52 –0.4 dB (80)________ 0.4 dB ±0.080 dB
–56 –0.4 dB (81)________ 0.4 dB ±0.078 dB
–60 –0.4 dB (82)________ 0.4 dB ±0.084 dB
–64 –0.4 dB (83)________ 0.4 dB ±0.081 dB
–68 –0.4 dB (84)________ 0.4 dB ±0.080 dB
–72 –0.4 dB (85)________ 0.4 dB ±0.084 dB
–76 –0.4 dB (86)________ 0.4 dB ±0.083 dB
–80 –0.4 dB (87)________ 0.4 dB ±0.089 dB
Linear Fidelity,
Res BW 1 kHz
dB from Ref Level
–4 –2.0% (93)________ 2.0% ±0.064%
–8 –2.0% (94)________ 2.0% ±0.064%
–12 –2.0% (95)________ 2.0% ±0.064%
–16 –2.0% (96)________ 2.0% ±0.064%
Table 3-2 Agilent E4401B Performance Verification Test Record
Agilent Technologies
Model E4401B Report No. ___________
Serial No. ___________ Date ___________
Test Description Minimum Results
Measured Maximum Measurement
Uncertainty