Chapter 2 271
Performance Verification Tests
31. Spurious Responses: Agilent E4402B and E4403B
5. Adjust the synthesized sweeper power level until the power meter 
reading is –10 dBm ±0.2 dB.
6. On the analyzer, press the following:
Peak Search (or Search), Marker, Delta 
FREQUENCY, 600 MHz
BW/Avg, 10
Wait for the “Vavg 10” to appear along the left side of the display.
7. On the analyzer, press Peak Search (or Search). The marker delta (∆ 
Mkr1) amplitude is the second harmonic suppression.
8. On the synthesized sweeper, press Marker, Off.
9. Calculate the 300 MHz Second Harmonic Intercept (SHI) using the 
second harmonic suppression value read in step 7 and the 300 MHz 
Frequency Response Error (FRE) from Table 2-73 as follows:
For example, if the second harmonic suppression is –59 dB, and the 
300 MHz FRE is –0.60 dB, the SHI would be 38.4 dBm:
10.Record the 300 MHz SHI as Entry 3 in the performance verification 
test record. 
Measuring 900 MHz Second Harmonic Distortion
1. Replace the 300 MHz low pass filter with the 1 GHz low pass filter 
as shown in Figure 2-48. 
2. On the synthesized sweeper, press the following:
CW, 900 MHz
POWER LEVEL, –10 dBm 
3. Enter the power sensor 1 GHz calibration factor into the power 
meter.
4. On the analyzer, press FREQUENCY, Center Freq, 900 MHz.
5. On the analyzer, press Peak Search (or Search).
6. Adjust the synthesized sweeper power level until the power meter 
reading is –10 dBm ±0.1 dB.
7. On the analyzer, press the following:
Peak Search (or Search)
Marker, Delta
FREQUENCY, Center Freq, 1.8 GHz
8. On the analyzer, press the following:
300 MHz  SHI 20 dBm– Second Harmonic Suppression 300 MHz FRE+–=