462 Chapter  3 
Performance Verification Test Records
Agilent E4402B Performance Verification Test Record
–10 dBm –0.3 dB (25)________ 0.3 dB ±0.091 dB
 0 dBm –0.3 dB (26)________ 0.3 dB ±0.100 dB
–30 dBm –0.3 dB (27)________ 0.3 dB ±0.092 dB
–40 dBm –0.3 dB (28)________ 0.3 dB ±0.094 dB
–50 dBm –0.5 dB (29)________ 0.5 dB ±0.095 dB
–60 dBm –0.5 dB (30)________ 0.5 dB ±0.097 dB
–70 dBm –0.5 dB (31)________ 0.5 dB ±0.099 dB
–80 dBm –0.7 dB (32)________ 0.7 dB ±0.101 dB
16. Resolution Bandwidth 
Switching Uncertainty
Resolution Bandwidth
 3 kHz –0.3 dB (1)________ 0.3 dB ±0.048 dB
 9 kHz –0.3 dB (2)________ 0.3 dB ±0.048 dB
 10 kHz –0.3 dB (3)________ 0.3 dB ±0.048 dB
 30 kHz –0.3 dB (4)________ 0.3 dB ±0.048 dB
 100 kHz –0.3 dB (5)________ 0.3 dB ±0.048 dB
 120 kHz –0.3 dB (6)________ 0.3 dB ±0.048 dB
 300 kHz –0.3 dB (7)________ 0.3 dB ±0.048 dB
 1 MHz –0.3 dB (8)________ 0.3 dB ±0.048 dB
 3 MHz –0.3 dB (9)________ 0.3 dB ±0.048 dB
 5 MHz –0.6 dB (10)________ 0.6 dB ±0.076 dB
 300 Hz (Option 1DR only) –0.3 dB (11)________ 0.3 dB ±0.048 dB
 200 Hz (Option 1DR only) –0.3 dB (12)________ 0.3 dB ±0.048 dB
 100 Hz (Option 1DR only) –0.3 dB (13)________ 0.3 dB ±0.048 dB
 30 Hz (Option 1DR only) –0.3 dB (14)________ 0.3 dB ±0.048 dB
 10 Hz (Option 1DR only) –0.3 dB (15)________ 0.3 dB ±0.048 dB
Table 3-4  Agilent E4402B Performance Verification Test Record
Agilent Technologies
Model E4402B Report No. ___________ 
Serial No. ___________ Date ___________
Test Description Minimum Results 
Measured Maximum Measurement 
Uncertainty