592 Chapter 3
Performance Verification Test Records
Agilent E4407B Performance Verification Test Record
6.9 GHz 13.0 dBm (13)________ 17.5 dBm ±0.16 dB
7.1 GHz 13.0 dBm (14)________ 17.5 dBm ±0.16 dB
0 to 55° C
First LO Frequency
2.9 GHz 15.0 dBm (1)________ 17.5 dBm ±0.16 dB
3.3 GHz 15.0 dBm (2)________ 17.5 dBm ±0.16 dB
3.7 GHz 15.0 dBm (3)________ 17.5 dBm ±0.16 dB
3.9 GHz 15.0 dBm (4)________ 17.5 dBm ±0.16 dB
4.1 GHz 15.0 dBm (5)________ 17.5 dBm ±0.16 dB
4.5 GHz 15.0 dBm (6)________ 17.5 dBm ±0.16 dB
4.9 GHz 15.0 dBm (7)________ 17.5 dBm ±0.16 dB
5.3 GHz 15.0 dBm (8)________ 17.5 dBm ±0.16 dB
5.7 GHz 15.0 dBm (9)________ 17.5 dBm ±0.16 dB
5.9 GHz 15.0 dBm (10)________ 17.5 dBm ±0.16 dB
6.1 GHz 15.0 dBm (11)________ 17.5 dBm ±0.16 dB
6.5 GHz 13.0 dBm (12)________ 17.5 dBm ±0.16 dB
6.9 GHz 13.0 dBm (13)________ 17.5 dBm ±0.16 dB
7.1 GHz 13.0 dBm (14)________ 17.5 dBm ±0.16 dB
53. IF Input Accuracy
(Option AYZ only)
Note: Enter data in the appropriate section below depending upon the
ambient temperature at which the test was performed.
20 to 30° C
IF Input Accuracy –1.0 dB (1)________ 1.0 dBm ±0.102 dB
0 to 50° C
IF Input Accuracy –1.5 dB (1)________ 1.5 dBm ±0.102 dB
Table 3-12 Agilent E4407B Performance Verification Test Record
Agilent Technologies
Model E4407B Report No. ___________
Serial No. ___________ Date ___________
Test Description Minimum Results
Measured Maximum Measurement
Uncertainty