Chapter  3 447
Performance Verification Test Records
Agilent E4401B Performance Verification Test Record
        1 MHz to 10 MHz (33)________ –99 dBmV ±1.23 dB
        10 MHz to 500 MHz (34)________ –100 dBmV ±1.23 dB
        500 MHz to 1 GHz (35)________ –100 dBmV ±1.23 dB
        1 GHz to 1.5 GHz (36)________ –94 dBmV ±1.23 dB
39. Residual Responses Note: Enter data in the appropriate section below depending upon the 
input impedance and serial number of the analyzer.
50 Ω, 150 kHz to 1.5 GHz (1)________ –90 dBm ±0.87 dB
75 Ω, 1 MHz to 1.5 GHz   (1)________ –36 dBmV ±0.87 dB
40. Fast Time Domain 
Amplitude Accuracy 
(Option AYX only)
 Amplitude Error –0.3% (1)________ 0.3% ±0.029%
41. Tracking Generator 
Absolute Amplitude and 
Vernier Accuracy
Note: Enter data in the appropriate section below depending upon the 
input impedance and serial number of the analyzer.
 (Option 1DN or Option 
1DQ only)
Absolute Amplitude 
Accuracy
–0.5 dB (1)________ 0.5 dB ±0.096 dB
Positive Vernier Accuracy (2)________ 0.75 dB ±0.008 dB
Negative Vernier 
Accuracy
–0.75 dB (3)________ ±0.008 dB
Power Sweep Accuracy (4)________ 1.5 dB ±0.008 dB
75 Ω (Option 1DQ)
Absolute Amplitude 
Accuracy
–1.5 dB (1)________ 1.5 dB ±0.096 dB
Positive Vernier Accuracy (2)________ 0.9 dB ±0.008 dB
Negative Vernier 
Accuracy
–0.9 dB (3)________ ±0.008 dB
Table 3-2  Agilent E4401B Performance Verification Test Record
Agilent Technologies
Model E4401B Report No. ___________ 
Serial No. ___________ Date ___________
Test Description Minimum Results 
Measured Maximum Measurement 
Uncertainty