446 Chapter 3
Performance Verification Test Records
Agilent E4401B Performance Verification Test Record
50 , 10 Hz RBW,
Preamp On
400 kHz (16)________ –150 dBm ±1.23 dB
1 MHz to 10 MHz (17)________ –150 dBm ±1.23 dB
10 MHz to 500 MHz (18)________ –154 dBm ±1.23 dB
500 MHz to 1 GHz (19)________ –152 dBm ±1.23 dB
1 GHz to 1.5 GHz (20)________ –150 dBm ±1.23 dB
75 , 1 kHz RBW,
Preamp Off
1 MHz to 10 MHz (21)________ –64 dBmV ±1.23 dB
10 MHz to 500 MHz (22)________ –66 dBmV ±1.23 dB
500 MHz to 1 GHz (23)________ –60 dBmV ±1.23 dB
1 GHz to 1.5 GHz (24)________ –56 dBmV ±1.23 dB
75 , 1 kHz RBW,
Preamp On:
1 MHz to 10 MHz (25)________ –80 dBmV ±1.23 dB
10 MHz to 500 MHz (26)________ –81 dBmV ±1.23 dB
500 MHz to 1 GHz (27)________ –81 dBmV ±1.23 dB
1 GHz to 1.5 GHz (28)________ –75 dBmV ±1.23 dB
75 , 10 Hz RBW,
Preamp Off
1 MHz to 10 MHz (29)________ –83 dBmV ±1.23 dB
10 MHz to 500 MHz (30)________ –85 dBmV ±1.23 dB
500 MHz to 1 GHz (31)________ –79 dBmV ±1.23 dB
1 GHz to 1.5 GHz (32)________ –75 dBmV ±1.23 dB
75 , 10 Hz RBW,
Preamp On:
Table 3-2 Agilent E4401B Performance Verification Test Record
Agilent Technologies
Model E4401B Report No. ___________
Serial No. ___________ Date ___________
Test Description Minimum Results
Measured Maximum Measurement
Uncertainty