
368 Chapter 2 
Performance Verification Tests
47. Tracking Generator Non-Harmonic Spurious Outputs: Agilent E4401B and E4411B (Option 
1DN or 1DQ)
analyzer to the next highest signal by pressing NEXT PEAK. Repeat 
step 4 above.
Perform step 7 only if the marker signal is a true response and not a 
fundamental or harmonic of the fundamental. Otherwise, continue 
with step 8.
7. Calculate the difference between the amplitude of marked signal 
and the fundamental amplitude as listed in Table 2-101.
For example, if the fundamental amplitude for a fundamental 
frequency of 10 MHz is 1.2 dBm and the marker amplitude is 
−40.8 dBm, the difference is −42 dBc.
Record this difference as the non-harmonic response amplitude for 
the appropriate analyzer center frequency and microwave analyzer 
start and stop frequency settings in Table 2-102. 
8. If a true non-harmonic spurious response is not found, record 
“NOISE” as the Amplitude of Non-Harmonic Response in Table 
2-102 for the appropriate analyzer center frequency and microwave 
analyzer start and stop frequency settings.
9. Repeat step 1 through step 8 for the remaining analyzer center 
frequency and microwave analyzer settings in Table 2-102.
Non-harmonic Amplitude Marker Amplitude Fundamental Amplitude–=
Table 2-102  1.5 GHz Tracking Generator Non-Harmonic Spurious Response 
Worksheet
 Analyzer 
Center 
Frequency
Microwave 
Analyzer Start 
Frequency
Microwave 
Analyzer Stop 
Frequency
Microwave 
Analyzer 
Resolution 
Bandwidth
Amplitude of 
Non-Harmonic 
Response
(dBc)
10 MHz 9 kHza100 kHza300 Hza
10 MHz 100 kHzb5 MHz 10 kHz
10 MHz 5 MHz 55 MHz 100 kHz
10 MHz 55 MHz 1240 MHz 1 MHz
10 MHz 1240 MHz 1500 MHz 1 MHz
750 MHz 9 kHza100 kHza300 Hza
750 MHz 100 kHzb5 MHz 10 kHz
750 MHz 5 MHz 55 MHz 100 kHz
750 MHz 55 MHz 1240 MHz 1 MHz
750 MHz 1240 MHz 1500 MHz 1 MHz
1.5 GHz 9 kHza100 kHza300 Hza