Chapter 2 389
Performance Verification Tests
52. First LO OUTPUT Power Accuracy (Option AYZ only)

Post-Test Instrument Restoration

12.Disconnect the power sensor from the 1st LO OUTPUT connector.

13.Replace the 50 termination on the 1st LO OUTPUT connector.

14.Press Preset the analyzer.

15.Press System, Alignments, Auto Align, All.

Table 2-107 First LO Output Amplitude Accuracy Worksheet

First LO
Frequency
(GHz)
Center
Frequency
(GHz)
Calibration
Factor
Frequency
(GHz)
Test Record
Entry
2.9 29 3.0 1)
3.3 33 3.0 2)
3.7 37 4.0 3)
3.9 39 4.0 4)
4.1 41 4.0 5)
4.5 45 5.0 6)
4.9 49 5.0 7)
5.3 53 5.0 8)
5.7 57 6.0 9)
5.9 59 6.0 10)
6.1 61 6.0 11)
6.5 65 7.0 12)
6.9 69 7.0 13)
7.1 71 7.0 14)