170 Chapter 2 
Performance Verification Tests
22. Frequency Response: Agilent E4401B and E4411B
10.On the analyzer, press Peak Search (or Search).
11.Adjust the function generator amplitude until the marker delta (∆ 
Mkr1) amplitude reads 0 dB ±0.05 dB.
12.Record the DVM reading in Table 2-42 as the DVM amplitude 
reading. 
13.Repeat step 8 through step 12 for each frequency in Table 2-42. 
14.For each of the frequencies in Table 2-42, subtract the DVM 
amplitude from the DVM Amplitude at 100 kHz recorded in step 6. 
Record the result as the Response Relative to 100 kHz in Table 2-42. 
15.From Table 2-41, note the Flatness Relative to 50 MHz for the 
100 kHz frequency. Record this below as the 100 kHz error relative 
to 50 MHz:
100 kHz Error Relative to 50 MHz __________ dB
16.Add the 100 kHz error relative to 50 MHz that was recorded in 
step 15 above to each of the Response Relative to 100 kHz entries in 
Table 2-42. Record the results as the Response Relative to 50 MHz in 
Table 2-42.
Test Results 
1. Enter the most positive number from the Flatness Relative to 
50 MHz column of Table 2-41: 
 ____________ dB 
2. Enter the most positive number from the Response Relative to 
50 MHz column of Table 2-42: 
 ____________ dB 
75 Ω inputs, Option 1DP: The frequency range below 100 kHz was not 
tested; no entry from Table 2-42 is necessary.
3. Record the more positive of numbers from step 1 and step 2 in 
Table 2-43 as the Maximum Response for Band 0. 
Table 2-42  Frequency Response Worksheet, ≤100 kHz
Frequency DVM 
Amplitude
Response 
Relative to 
100 kHz
Response 
Relative to 
50 MHz
100 kHz 0 dB (Ref)
75 kHz
50 kHz
20 kHz
9 kHz