Chapter 2 347
Performance Verification Tests
43. Tracking Generator Level Flatness: Agilent E4401B and E4411B (Option 1DN or 1DQ)

Tracking Generator Level Flatness, Center

Frequency 100 kHz

NOTE Perform step 1 to step 7 for 50 tracking generators only
(Option 1DN).
1. Set up the digital multimeter as follows.
NOTE Perform step 2 to step 7 for each measurement value in Table 2-97.
2. Refer to Figure 2-62 to set up the equipment.
Figure 2-62 Tracking Generator Level Flatness Test Setup, 100 kHz
3. To set the analyzer center frequency to 9 kHz, press FREQUENCY,
9kHz (or as indicated in Table 2-96).
Parameter Setting
AC/DC AC Volts
Impedance & Units:
Set to 50 impedance SMATH 10a
Set to dBm MATH 5a
Set to Synchronous SETACV 3a
Sub-sampled mode
a. To set the Agilent 3458A multimeter functions from the
front panel, press the blue shift key, then Recall State (T)
key. Use the (up) and (down) arrows to select the
appropriate function, then enter the value from the
numeric keypad and press enter.