206 Chapter 2
Performance Verification Tests
24. Frequency Response, Agilent E4404B, E4405B, E4407B, and E4408B
50 MHz (dc) column of Table 2-52 for frequencies between 6.7 GHz
and 13.2 GHz. Record this number in Table 2-53 as the Minimum
Response for Band 2.
14.For Band 2 in Table 2-53, subtract the Minimum Response value
from the Maximum Response value and record the result in the
Peak-to-Peak Response column.
15.Record the most positive number from the Flatness Relative to
50 MHz (ac) column of Table 2-50 as the Maximum Response for
Band 0 in Table 2-54.
16.Record the most negative number from the Flatness Relative to
50 MHz (ac) column of Table 2-50 as the Minimum Response for
Band 0 in Table 2-54.
17.For Band 0 in Table 2-54, subtract the Minimum Response value
from the Maximum Response value and record the result in the
Peak-to-Peak Response column.
18.If the analyzer is an Agilent E4404B with a serial number less than
US39440498 or an E4405B with a serial number less than
US39440327, then continue with step 25.
19.Note the most positive number from the Flatness Relative to 50 MHz
(ac) column of Table 2-50 for frequencies between 800 MHz and
1.0 GHz. Record this number in Table 2-54 as the Maximum
Response for Band 0A.
20.Note the most negative number from the Flatness Relative to
50 MHz (ac) column of Table 2-50 for frequencies between 800 MHz
and 1.0 GHz. Record this number in Table 2-54 as the Minimum
Response for Band 0A.
21.For Band 0A in Table 2-54, subtract the Minimum Response value
from the Maximum Response value and record the result in the
Peak-to-Peak Response column.
22.Note the most positive number from the Flatness Relative to 50 MHz
(ac) column of Table 2-50 for frequencies between 1.7 GHz and
2.0 GHz. Record this number in Table 2-54 as the Maximum
Response for Band 0B.
23.Note the most negative number from the Flatness Relative to
50 MHz (ac) column of Table 2-50 for frequencies between 1.7 GHz
and 2.0 GHz. Record this number in Table 2-54 as the Minimum
Response for Band 0B.
24.For Band 0B in Table 2-54, subtract the Minimum Response value
from the Maximum Response value and record the result in the
Peak-to-Peak Response column.
25.Note the most positive number from the Flatness Relative to 50 MHz
(ac) column of Table 2-52 for frequencies between 3.0 GHz and